FDA Enforcement Class II Ongoing

Toco M2734A, that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341

Recall: Z-1165-2020 · Reported February 19, 2020

Enforcement

Recall Number
Z-1165-2020
Event ID
83610
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Pacific Medical Group Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 19, 2020
Initiation Date
September 11, 2019
Classification Date
February 12, 2020
Address
212 Avenida Fabricante, San Clemente, CA, 92672-7538, United States

Description

Toco M2734A, that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341

Reason

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.

Code Info

All serial numbers serviced between 9/2/15 and 4/2/19.

Distribution

Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR, ZAF, BHR, ESP, ITA, FRA, VUT, AUS, GBR.

Quantity

466