FDA Enforcement Class II Ongoing

Cardinal Health Presource Packs IV START KIT- Intended for IV access device CATALOG #: 01-5798B

Recall: Z-1157-2023 · Reported March 1, 2023

Enforcement

Recall Number
Z-1157-2023
Event ID
91560
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Cardinal Health 200, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
March 1, 2023
Initiation Date
November 18, 2022
Classification Date
February 22, 2023
Address
3651 Birchwood Dr, N/A, Waukegan, IL, 60085-8337, United States

Description

Cardinal Health Presource Packs IV START KIT- Intended for IV access device CATALOG #: 01-5798B

Reason

Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity

Code Info

UDI-DI: 50195594577153(cs) 10195594577155(ea) Lot Number: 816643 Exp. Date: 2/1/2025

Distribution

US Nationwide distribution in the states of GA, PA.

Quantity

4560 EA