FDA Enforcement
Class II
Terminated
Medicrea TLIF-B Holder, short intended for Intervertebral body fusion device Catalog: Number A24200010
Recall: Z-1156-2020
·
Reported February 19, 2020
Enforcement
- Recall Number
- Z-1156-2020
- Event ID
- 84770
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medicrea International
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 19, 2020
- Initiation Date
- October 2, 2019
- Classification Date
- February 12, 2020
- Termination Date
- February 8, 2021
- Address
- 5389 Route De Strasbourg, Vancia, Rillieux La Pape, N/A, France
Description
Medicrea TLIF-B Holder, short intended for Intervertebral body fusion device Catalog: Number A24200010
Reason
Inner draw rod, which threads into the cage may potentially be separated from the proximal knob of the inserter
Code Info
Lot Numbers l9B0570; 19B0572
Distribution
OH, VA
Quantity
14 units