FDA Enforcement Class II Terminated

Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.

Recall: Z-1156-2014 · Reported March 19, 2014

Enforcement

Recall Number
Z-1156-2014
Event ID
67495
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mindray DS USA, Inc. dba Mindray North America
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 19, 2014
Initiation Date
January 8, 2014
Classification Date
March 7, 2014
Termination Date
January 21, 2015
Address
800 MacArthur Blvd, N/A, Mahwah, NJ, 07430-2001, United States

Description

Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.

Reason

Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindray¿s V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.

Code Info

p/n 0998-00-0708-01; 0998-UC-0708-01 and 0020-00-0205-XXX

Distribution

Worldwide Distribution - USA Nationwide Distribution

Quantity

732 units in the US