Respironics V680 Ventilator, Service # 850011, (OUS Only)

Recall: Z-1155-2020 · Reported February 19, 2020

Enforcement

Recall Number
Z-1155-2020
Event ID
83980
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Respironics California, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
February 19, 2020
Initiation Date
September 11, 2019
Classification Date
February 12, 2020
Address
2271 Cosmos Ct, Carlsbad, CA, 92011-1517, United States

Description

Respironics V680 Ventilator, Service # 850011, (OUS Only)

Reason

Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.

Code Info

Serial Numbers: 100255874, 100255957, 100257487, 100257488, 100257490, 100257491, 100257493, 100257494, 100257495, 100257496, 100257497, 100257498, 100257499, 100257500, 100257501, 100258060, 100258061, 100258063, 100258064, 100258065, 100258066, 100258067, 100258068, 100260220, 100260221, 100260222, 100260223, 100260224, 100260227, 100260228, 100260229, 100260616, 100260617, 100260618, 100260619, 100260620, 100260621, 100260622, 100260623, 100260624, 100260625, 100260626, 100260627, 100260628, 100260629, 100260630, 100260631, 100260632, 100260633, 100260634, 100260635, 100261843, 100261844, 100261845, 100261846, 100261847, 100261848, 100261849, 100261850, 100261851, 100261852, 100261853, 100261854, 100261855, 100261856, 100261857, 100261900, 100261901, 100261902, 100263992, 100263993, 100263994, 100263995, 100263996, 100263997, 100263998, 100263999, 100264020, 100264021, 100264022, 100264023, 100264024, 100264025, 100264026, 100264279, 100264300, 100264301, 100264302, 100264303, 100264561, 100264562, 100264563, 100264564, 100264565, 100264566, 100264567, 100264568, 100264569, 100264570, 100264571, 100264572, 100264573, 100264574, 100264575, 100264576, 100264577, 100264578, 100264579, 100264780, 100264781, 100264782, 100264783, 100264784, 100264785, 100265624, 100265625, 100265626, 100265627, 100265628, 100265629, 100265630, 100265631, 100265632, 100265633, 100265634, 100265635, 100265636, 100265637, 100265638, 100265639, 100265640, 100265641, 100265642, 201034917, 201034918, 201034919, 201034945, 201034946, 201034947, 201034949, 201034950, 201034951, 201034952, 201034953, 201034954, 201035015, 201035016, 201035017, 201035018, 201035019, 201035080, 201035081, 201035082, 201035083, 201035084, 201035085, 201035197, 201035198, 201035199, 201035221, 201035222, 201035223, 201035224, 201035225, 201035226, 201035320, 201035321, 201035322, 201035323, 201035324, 201035325, 201035326, 201035328, 201035329, 201035330, 201035331, 201035332, 201035333, 201035334, 201035335, 201035336, 201035337, 201035338, 201035339, 201035360, 201035361, 201035362, 201035363, 201035364, 201035365

Distribution

US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam

Quantity

190