Respironics RP-UI Assy, V60 Ventilators User Interface RP-Kits, REF/P/N: 453561528931, 453561528941, 453561536451, 453561536461
Enforcement
- Recall Number
- Z-1153-2020
- Event ID
- 83980
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Respironics California, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- February 19, 2020
- Initiation Date
- September 11, 2019
- Classification Date
- February 12, 2020
- Address
- 2271 Cosmos Ct, Carlsbad, CA, 92011-1517, United States
Description
Respironics RP-UI Assy, V60 Ventilators User Interface RP-Kits, REF/P/N: 453561528931, 453561528941, 453561536451, 453561536461
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.
Serial Numbers: 100270400, 100270407, 100256220, 100270404, 100263004, 100266507, 100270402, 100263001, 100263009, 100258193, 100258197, 100258883, 100262978, 100263006, 100258196, 100256196, 100256197, 100256198, 100263003, 100266499, 100266505, 100258889, 100256192, 100258887, 100266509, 100270408, 100262976, 100258881, 100258880, 100266511, 100270406, 100258885, 100256223, 100263000, 100258884, 100258888, 100266500, 100266504, 100258886, 100256222, 100256195, 100256199, 100266502, 100266508, 100266510, 100270403, 100270490, 100257708, 100257713, 100257716, 100256831, 100256834, 100256837, 100262373, 100262556, 100262575, 100265136
US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam
56