Respironics V60 Ventilator, Service # 850008, REF: 1053613, 1053614, 1053617, 1053614
Enforcement
- Recall Number
- Z-1152-2020
- Event ID
- 83980
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Respironics California, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- February 19, 2020
- Initiation Date
- September 11, 2019
- Classification Date
- February 12, 2020
- Address
- 2271 Cosmos Ct, Carlsbad, CA, 92011-1517, United States
Description
Respironics V60 Ventilator, Service # 850008, REF: 1053613, 1053614, 1053617, 1053614
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchscreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.
Serial Numbers: 100005249 100254792 100255723 100255865 100256311 100256502 100256609 100256781 100256875 100257173 100257312 100257583 100257698 100258034 100258360 100258513 100258809 100259082 100259249 100259633 100259878 100260186 100260534 100260846 100261115 100261278 100261538 100261743 100262170 100262417 100262531 100262627 100262821 100007779 100254793 100255724 100255866 100256312 100256503 100256610 100256782 100256876 100257174 100257313 100257584 100257699 100258035 100258361 100258514 100258810 100259083 100259250 100259634 100259879 100260231 100260535 100260847 100261117 100261279 100261539 100261744 100262171 100262418 100262532 100262628 100262822 100007982 100254977 100255725 100255965 100256313 100256504 100256611 100256783 100256877 100257175 100257314 100257585 100257800 100258159 100258362 100258553 100258811 100259085 100259251 100259635 100259880 100260233 100260537 100260849 100261120 100261302 100261540 100261745 100262172 100262419 100262533 100262629 100262823 100008795 100255098 100255726 100255966 100256314 100256505 100256612 100256784 100256980 100257176 100257316 100257586 100257801 100258211 100258363 100258554 100258812 100259086 100259252 100259636 100259881 100260234 100260538 100260850 100261121 100261320 100261541 100261746 100262173 100262420 100262534 100262630 100262824 100009556 100255165 100255727 100255967 100256315 100256507 100256613 100256785 100256981 100257177 100257317 100257587 100257802 100258240 100258364 100258556 100258813 100259087 100259253 100259724 100259882 100260235 100260539 100260851 100261122 100261321 100261543 100261747 100262174 100262421 100262535 100262631 100262825 100010513 100255317 100255728 100255968 100256316 100256508 100256614 100256786 100256982 100257178 100257318 100257588 100257803 100258241 100258365 100258557 100258814 100259088 100259255 100259726 100259996 100260236 100260540 100260852 100261123 100261322 100261544 100261748 100262175 100262422 100262536 100262632 100262826 100011641 100255318 100255729 100255969 100256317 100256509 100256615 100256842 100256983 100257179 100257319 100257589 100257804 100258242 100258366 100258558 100258817 100259089 100259256 100259732 100259997 100260237 100260541 100260853 100261124 100261365 100261545 100261749 100262176 100262423 100262537 100262633 100262827 100015789 100255409 100255730 100255970 100256318 100256510 100256616 100256843 100256984 100257200 100257377 100257590 100257805 100258243 100258367 100258559 100258818 100259090 100259258 100259735 100259998 100260238 100260542 100260854 100261125 100261366 100261546 100261750 100262177 100262424 100262539 100262634 100262828 100020831 100255421 100255732 100255971 100256319 100256511 100256617 100256844 100256985 100257201 100257378 100257591 100257806 100258244 100258368 100258587 100258819 100259091 100259259 100259737 100259999 100260239 100260545 100260855 100261126 100261367 100261547 100261751 100262178 100262426 100262540 100262635 100262829 100021459 100255422 100255733 100255972 100256395 100256512 100256618 100256845 100256986 100257202 100257379 100257592 100257807 100258245 100258450 100258588 100258820 100259092 100259260 100259783 100260020 100260240 100260546 100260856 100261127 100261368 100261549 100261752 100262179 100262427 100262541 100262637 100262830 100021751 100255424 100255734 100255973 100256396 100256513 100256619 100256846 100256987 100257203 100257380 100257593 100257808 100258246 100258451 100258589 100258821 100259093 100259261 100259784 100260021 100260241 100260547 100260857 100261128 100261369 100261550 100261753 100262180 100262429 100262543 100262638 100262832 100023634 100255500 100255737 100255974 100256398 100256514 100256620 100256847 100256988 100257204 100257381 100257670 100257809 100258247 100258452 100258590 100258822 100259094 100259262 100259785 100260022 100260242 100260548 100260858 100261129 100261370 100261551 100261754 100262181 100262431 100262544 100262639 100262833 100023729 100255501 100255738 100255975 100256399 100256515 100256621 100256848 100256990 100257205 100257382 100257671 100257811 100258248 100258453 100258591 100258823 100259095 100259263 100259786 100260024 100260243 100260549 100260951 100261130 100261371 100261552 100261755 100262182 100262432 100262546 100262640 100262834 100026949 100255502 100255739 100255976 100256400 100256516 100256623 100256849 100256991 100257206 100257383 100257672 100257812 100258249 100258454 100258592 100258920 100259096 100259380 100259788 100260025 100260244 100260550 100260953 100261131 100261372 100261553 100261756 100262183 100262433 100262547 100262641 100262835 100029508 100255504 100255840 100256005 100256401 100256517 100256624 100256850 100257085 100257207 100257384 100257673 100257813 100258250 100258455 100258593 100258921 100259097 100259381 100259791 100260026 100260245 100260551 100260955 100261132 100261373 100261554 100261757 100262184 100262434 100262548 100262642 100262836 100034162 100255506 100255841 100256127 100256402 100256518 100256625 100256851 100257086 100257208 100257385 100257674 100257814 100258272 100258456 100258594 100258922 100259098 100259382 100259792 100260027 100260246 100260552 100260956 100261133 100261374 100261555 100261758 100262185 100262436 100262549 100262750 100262837 100034395 100255507 100255842 100256128 100256403 100256519 100256626 100256852 100257087 100257209 100257386 100257675 100257815 100258273 100258457 100258596 100258923 100259099 100259383 100259794 100260028 100260247 100260553 100260957 100261134 100261375 100261556 100261759 100262186 100262437 100262550 100262751 100262839 100066214 100255509 100255843 100256129 100256404 100256520 100256627 100256853 100257089 100257210 100257387 100257676 100257816 100258274 100258471 100258597 100258924 100259100 100259384 100259796 100260030 100260248 100260554 100260958 100261135 100261376 100261557 100262080 100262187 100262438 100262551 100262753 100262940 100082340 100255510 100255844 100256130 100256405 100256521 100256628 100256854 100257090 100257211 100257388 100257677 100257817 100258275 100258472 100258598 100258925 100259102 100259385 100259797 100260031 100260249 100260708 100260960 100261136 100261378 100261558 100262082 100262188 100262439 100262552 100262754 100262942 100246732 100255512 100255845 100256131 100256406 100256522 100256629 100256855 100257091 100257214 100257389 100257678 100257818 100258276 100258473 100258600 100258926 100259103 100259386 100259853 100260032 100260251 100260710 100260961 100261137 100261379 100261559 100262083 100262189 100262440 100262603 100262755 100262943 100248301 100255513 100255846 100256132 100256407 100256523 100256630 100256856 100257092 100257215 100257390 100257679 100257819 100258277 100258474 100258601 100258927 100259135 100259387 100259854 100260033 100260252 100260711 100260962 100261138 100261400 100261560 100262084 100262190 100262441 100262604 100262756 100262944 100250369 100255547 100255847 100256133 100256408 100256563 100256631 100256857 100257093 100257216 100257391 100257680 100257921 100258278 100258475 100258602 100258928 100259136 100259388 100259858 100260034 100260253 100260712 100260963 100261139 100261401 100261561 100262085 100262191 100262507 100262605 100262757 100262945 100252951 100255548 100255848 100256134 100256410 100256564 100256632 100256858 100257094 100257217 100257392 100257681 100257922 100258279 100258476 100258603 100258929 100259137 100259389 100259860 100260035 100260254 100260713 100260964 100261140 100261402 100261562 100262086 100262192 100262508 100262606 100262758 100262946 100254466 100255556 100255849 100256135 100256411 100256565 100256633 100256859 100257095 100257218 100257393 100257682 100257923 100258280 100258477 100258604 100258930 100259138 100259390 100259862 100260036 100260255 100260714 100260965 100261168 100261404 100261563 100262087 100262193 100262510 100262608 100262759 100262947 100254552 100255558 100255850 100256136 100256412 100256566 100256634 100256860 100257096 100257219 100257394 100257683 100257924 100258281 100258478 100258606 100258931 100259139 100259391 100259863 100260037 100260256 100260715 100260966 100261169 100261406 100261564 100262088 100262194 100262511 100262610 100262805 100262949 100254738 100255594 100255851 100256137 100256413 100256567 100256635 100256861 100257097 100257260 100257395 100257684 100257925 100258282 100258479 100258607 100258932 100259234 100259392 100259864 100260038 100260257 100260831 100260967 100261170 100261407 100261667 100262089 100262195 100262512 100262611 100262806 100262950 100254739 100255596 100255852 100256138 100256415 100256568 100256636 100256862 100257098 100257261 100257396 100257685 100258020 100258283 100258495 100258609 100258933 100259235 100259393 100259865 100260039 100260259 100260832 100260968 100261171 100261408 100261668 100262090 100262196 100262513 100262612 100262807 100262951 100254745 100255597 100255853 100256139 100256416 100256569 100256637 100256863 100257099 100257262 100257397 100257686 100258021 100258284 100258496 100258610 100258934 100259236 100259394 100259866 100260040 100260440 100260833 100260970 100261172 100261409 100261669 100262131 100262198 100262514 100262613 100262808 100262952 100254746 100255598 100255854 100256300 100256417 100256570 100256638 100256864 100257162 100257263 100257398 100257687 100258022 100258285 100258497 100258611 100258935 100259237 100259395 100259867 100260042 100260451 100260834 100260971 100261173 100261410 100261670 100262132 100262199 100262515 100262616 100262809 100262953 100254747 100255599 100255855 100256301 100256418 100256571 100256639 100256865 100257163 100257264 100257399 100257688 100258023 100258286 100258499 100258613 100258936 100259238 100259396 100259868 100260043 100260452 100260835 100260972 100261174 100261411 100261671 100262133 100262200 100262520 100262617 100262810 100262954 100254748 100255620 100255856 100256302 100256419 100256572 100256640 100256866 100257164 100257265 100257400 100257689 100258024 100258289 100258500 100258614 100258937 100259239 100259397 100259869 100260044 100260525 100260836 100260973 100261175 100261412 100261672 100262134 100262201 100262522 100262618 100262811 100262955 100254749 100255621 100255857 100256303 100256420 100256573 100256641 100256867 100257165 100257266 100257401 100257690 100258025 100258290 100258501 100258800 100258938 100259240 100259398 100259870 100260045 100260526 100260837 100260974 100261176 100261413 100261673 100262135 100262202 100262523 100262619 100262812 100262956 100254750 100255622 100255858 100256304 100256422 100256574 100256642 100256868 100257166 100257267 100257403 100257691 100258026 100258291 100258505 100258801 100258939 100259241 100259399 100259871 100260174 100260527 100260838 100261002 100261177 100261414 100261674 100262136 100262203 100262524 100262620 100262813 100262957 100254751 100255623 100255859 100256305 100256423 100256575 100256643 100256869 100257167 100257268 100257404 100257692 100258027 100258292 100258506 100258802 100258941 100259242 100259400 100259872 100260175 100260528 100260839 100261003 100261178 100261452 100261675 100262137 100262204 100262525 100262621 100262814 100262958 100254752 100255626 100255860 100256306 100256424 100256576 100256644 100256870 100257168 100257269 100257405 100257693 100258028 100258293 100258507 100258803 100258942 100259243 100259401 100259873 100260177 100260529 100260840 100261004 100261179 100261453 100261676 100262138 100262205 100262526 100262622 100262815 100262959 100254783 100255676 100255861 100256307 100256425 100256577 100256645 100256871 100257169 100257270 100257406 100257694 100258029 100258295 100258508 100258805 100259056 100259244 100259402 100259874 100260179 100260530 100260841 100261111 100261273 100261454 100261677 100262139 100262206 100262527 100262623 100262816 100262960 100254785 100255678 100255862 100256308 100256426 100256578 100256646 100256872 100257170 100257271 100257580 100257695 100258030 100258297 100258509 100258806 100259057 100259245 100259403 100259875 100260180 100260531 100260842 100261112 100261274 100261455 100261678 100262167 100262207 100262528 100262624 100262817 100262961 100254787 100255721 100255863 100256309 100256500 100256607 100256647 100256873 100257171 100257310 100257581 100257696 100258031 100258298 100258510 100258807 100259059 100259246 100259404 100259876 100260182 100260532 100260843 100261113 100261275 100261456 100261679 100262168 100262209 100262529 100262625 100262818 100262962 100254790 100255722 100255864 100256310 100256501 100256608 100256780 100256874 100257172 100257311 100257582 100257697 100258032 100258299 100258511 100258808 100259081 100259248 100259632 100259877 100260183 100260533 100260845 100261114 100261276 100261537 100261742 100262169 100262210 100262530 100262626 100262819 100262963 100262964 100263281 100263541 100263659 100263704 100264442 100265434 100265700 100266018 100266156 100266267 100266523 100266687 100266874 100267094 100267311 100267441 100267573 100267848 100267989 100268193 100268332 100268500 100268626 100268894 100269107 100269155 100269326 100269611 100269781 100269892 100270033 100270287 100262965 100263282 100263542 100263660 100263705 100264443 100265435 100265701 100266020 100266157 100266369 100266524 100266688 100266875 100267095 100267312 100267442 100267574 100267849 100267990 100268194 100268333 100268501 100268627 100268895 100269108 100269156 100269327 100269612 100269782 100269893 100270034 100270288 100262966 100263283 100263543 100263661 100263706 100264444 100265436 100265702 100266021 100266158 100266370 100266525 100266689 100266876 100267096 100267313 100267443 100267688 100267850 100267991 100268195 100268334 100268502 100268628 100268896 100269109 100269157 100269328 100269613 100269783 100269894 100270035 100270289 100262967 100263284 100263544 100263662 100263708 100264445 100265437 100265703 100266022 100266159 100266371 100266526 100266690 100266877 100267097 100267314 100267444 100267689 100267851 100267992 100268244 100268335 100268503 100268629 100268897 100269110 100269158 100269329 100269614 100269784 100269895 100270036 100270290 100262968 100263285 100263545 100263663 100263709 100264581 100265438 100265704 100266023 100266160 100266372 100266527 100266691 100266878 100267098 100267315 100267445 100267690 100267852 100267993 100268245 100268337 100268504 100268630 100268911 100269111 100269159 100269330 100269615 100269785 100269896 100270037 100270291 100262969 100263286 100263546 100263664 100264295 100264582 100265439 100265705 100266024 100266161 100266374 100266528 100266692 100266879 100267099 100267316 100267446 100267691 100267853 100267994 100268246 100268338 100268505 100268631 100268912 100269112 100269160 100269331 100269616 100269786 100269897 100270038 100270292 100262970 100263287 100263547 100263665 100264297 100264583 100265459 100265706 100266025 100266163 100266375 100266529 100266693 100266900 100267120 100267317 100267447 100267692 100267854 100267995 100268247 100268339 100268506 100268860 100268913 100269113 100269161 100269332 100269617 100269787 100269898 100270039 100270293 100262971 100263288 100263548 100263666 100264298 100264584 100265460 100265708 100266026 100266164 100266376 100266530 100266694 100266901 100267121 100267318 100267448 100267693 100267855 100267996 100268248 100268426 100268507 100268861 100268914 100269115 100269162 100269333 100269618 100269788 100269899 100270040 100270294 100262972 100263289 100263549 100263667 100264299 100264585 100265461 100265820 100266027 100266165 100266377 100266531 100266742 100266902 100267122 100267319 100267449 100267694 100267856 100267997 100268249 100268427 100268508 100268862 100268915 100269116 100269163 100269334 100269619 100269789 100270001 100270252 100270295 100262973 100263290 100263550 100263668 100264406 100264586 100265462 100265821 100266028 100266166 100266378 100266532 100266743 100266903 100267123 100267320 100267450 100267695 100267857 100267998 100268250 100268428 100268509 100268863 100268916 100269117 100269164 100269335 100269621 100269790 100270002 100270254 100270296 100262975 100263291 100263551 100263669 100264407 100264587 100265463 100265822 100266029 100266167 100266379 100266533 100266744 100266904 100267124 100267321 100267451 100267696 100267859 100268000 100268251 100268429 100268510 100268864 100268917 100269118 100269165 100269336 100269622 100269791 100270003 100270255 100270297 100263120 100263292 100263552 100263670 100264408 100264588 100265464 100265823 100266030 100266168 100266400 100266534 100266745 100266905 100267125 100267322 100267452 100267697 100267860 100268164 100268252 100268430 100268511 100268865 100268918 100269119 100269166 100269582 100269623 100269792 100270004 100270256 100270298 100263121 100263293 100263555 100263671 100264410 100264589 100265465 100265824 100266031 100266198 100266401 100266535 100266746 100266906 100267126 100267323 100267453 100267698 100267861 100268165 100268253 100268431 100268512 100268866 100268919 100269127 100269167 100269583 100269624 100269793 100270005 100270258 100270299 100263122 100263294 100263556 100263672 100264411 100264590 100265466 100265825 100266033 100266199 100266402 100266536 100266747 100266907 100267127 100267324 100267525 100267699 100267862 100268166 100268254 100268432 100268513 100268867 100269080 100269128 100269168 100269584 100269625 100269794 100270006 100270259 100270300 100263123 100263295 100263557 100263673 100264412 100264591 100265467 100265826 100266034 100266240 100266403 100266537 100266748 100266908 100267128 100267325 100267526 100267700 100267863 100268167 100268255 100268433 100268600 100268868 100269081 100269129 100269169 100269585 100269626 100269795 100270007 100270260 100270301 100263124 100263296 100263558 100263674 100264413 100264592 100265468 100265827 100266035 100266241 100266404 100266538 100266749 100266909 100267129 100267326 100267527 100267701 100267864 100268168 100268256 100268434 100268601 100268869 100269082 100269130 100269170 100269586 100269627 100269796 100270008 100270261 100270460 100263125 100263297 100263559 100263675 100264414 100264593 100265469 100265828 100266036 100266242 100266405 100266539 100266750 100266910 100267130 100267327 100267528 100267702 100267865 100268169 100268258 100268435 100268602 100268870 100269083 100269131 100269300 100269587 100269628 100269797 100270009 100270262 100270461 100263126 100263298 100263561 100263676 100264415 100264595 100265470 100265829 100266037 100266243 100266406 100266540 100266751 100266911 100267131 100267328 100267529 100267703 100267966 100268170 100268259 100268436 100268603 100268871 100269084 100269132 100269301 100269588 100269629 100269798 100270010 100270263 100270462 100263128 100263299 100263562 100263677 100264416 100264596 100265471 100265831 100266038 100266244 100266407 100266541 100266752 100266912 100267132 100267419 100267531 100267704 100267967 100268171 100268263 100268437 100268604 100268872 100269085 100269133 100269302 100269589 100269630 100269799 100270011 100270264 100270463 100263129 100263300 100263563 100263678 100264417 100264597 100265472 100265832 100266039 100266245 100266408 100266542 100266753 100266913 100267133 100267420 100267532 100267705 100267968 100268172 100268264 100268438 100268605 100268873 100269086 100269134 100269303 100269590 100269631 100269800 100270012 100270265 100270464 100263130 100263301 100263564 100263680 100264418 100264598 100265679 100265834 100266040 100266246 100266409 100266619 100266754 100266914 100267134 100267421 100267533 100267706 100267969 100268173 100268265 100268439 100268606 100268874 100269087 100269135 100269304 100269591 100269632 100269801 100270013 100270266 100270465 100263131 100263302 100263565 100263682 100264420 100264599 100265680 100265835 100266041 100266247 100266410 100266665 100266755 100266915 100267135 100267422 100267534 100267707 100267970 100268174 100268266 100268480 100268607 100268875 100269088 100269136 100269305 100269592 100269633 100269802 100270014 100270267 100270466 100263132 100263303 100263566 100263683 100264421 100264600 100265682 100265836 100266042 100266248 100266411 100266666 100266756 100266916 100267136 100267423 100267535 100267708 100267971 100268175 100268267 100268481 100268608 100268876 100269089 100269137 100269306 100269593 100269634 100269803 100270015 100270268 100270467 100263133 100263304 100263567 100263684 100264423 100264601 100265683 100265837 100266043 100266249 100266412 100266667 100266757 100266917 100267137 100267424 100267536 100267811 100267972 100268176 100268268 100268482 100268609 100268877 100269090 100269138 100269307 100269594 100269635 100269804 100270016 100270269 100270468 100263134 100263305 100263568 100263685 100264424 100264602 100265684 100265838 100266140 100266250 100266413 100266668 100266758 100266918 100267138 100267425 100267537 100267812 100267973 100268177 100268269 100268483 100268610 100268878 100269091 100269139 100269308 100269595 100269636 100269805 100270017 100270270 100270469 100263135 100263306 100263569 100263686 100264425 100264603 100265685 100265839 100266141 100266251 100266414 100266669 100266759 100266919 100267139 100267426 100267538 100267813 100267974 100268178 100268270 100268485 100268611 100268879 100269092 100269140 100269309 100269596 100269637 100269806 100270018 100270271 100270470 100263136 100263307 100263570 100263687 100264426 100264604 100265686 100265840 100266142 100266252 100266415 100266670 100266760 100266920 100267140 100267427 100267539 100267814 100267975 100268179 100268271 100268486 100268612 100268880 100269093 100269141 100269310 100269597 100269638 100269807 100270019 100270272 100270471 100263137 100263308 100263571 100263688 100264427 100264605 100265687 100265841 100266143 100266253 100266416 100266672 100266761 100266921 100267141 100267428 100267560 100267815 100267976 100268180 100268272 100268487 100268613 100268881 100269094 100269142 100269311 100269598 100269768 100269808 100270020 100270273 100270472 100263139 100263309 100263572 100263689 100264428 100264606 100265688 100265842 100266144 100266254 100266417 100266673 100266762 100267082 100267142 100267429 100267561 100267836 100267977 100268181 100268320 100268488 100268614 100268882 100269095 100269143 100269312 100269599 100269769 100269809 100270021 100270274 100270473 100263189 100263310 100263573 100263690 100264429 100264607 100265689 100265843 100266145 100266255 100266418 100266674 100266763 100267083 100267300 100267430 100267562 100267837 100267978 100268182 100268321 100268489 100268615 100268883 100269096 100269144 100269313 100269600 100269770 100269810 100270022 100270275 100270474 100263190 100263311 100263574 100263691 100264431 100264608 100265690 100265844 100266146 100266257 100266513 100266675 100266764 100267084 100267301 100267431 100267563 100267838 100267979 100268183 100268322 100268490 100268616 100268884 100269097 100269145 100269314 100269601 100269771 100269811 100270023 100270277 100270475 100263191 100263312 100263575 100263692 100264432 100264609 100265691 100265845 100266147 100266258 100266514 100266676 100266765 100267085 100267302 100267432 100267564 100267839 100267980 100268184 100268323 100268491 100268617 100268885 100269098 100269146 100269315 100269602 100269772 100269812 100270024 100270278 100270476 100263193 100263313 100263576 100263693 100264433 100264610 100265692 100265846 100266148 100266259 100266515 100266677 100266766 100267086 100267303 100267433 100267565 100267840 100267981 100268185 100268324 100268492 100268618 100268886 100269099 100269147 100269316 100269603 100269773 100269813 100270025 100270279 100270477 100263194 100263315 100263577 100263694 100264434 100264611 100265693 100265847 100266149 100266260 100266516 100266678 100266767 100267087 100267304 100267434 100267566 100267841 100267982 100268186 100268325 100268493 100268619 100268887 100269100 100269148 100269317 100269604 100269774 100269814 100270026 100270280 100270478 100263195 100263535 100263578 100263696 100264435 100264612 100265694 100265848 100266150 100266261 100266517 100266679 100266768 100267088 100267305 100267435 100267567 100267842 100267983 100268187 100268326 100268494 100268620 100268888 100269101 100269149 100269318 100269605 100269775 100269815 100270027 100270281 100270479 100263196 100263536 100263579 100263697 100264436 100264613 100265695 100265849 100266151 100266262 100266518 100266680 100266769 100267089 100267306 100267436 100267568 100267843 100267984 100268188 100268327 100268495 100268621 100268889 100269102 100269150 100269319 100269606 100269776 100269816 100270028 100270282 100270480 100263197 100263537 100263580 100263698 100264437 100265429 100265696 100266014 100266152 100266263 100266519 100266682 100266770 100267090 100267307 100267437 100267569 100267844 100267985 100268189 100268328 100268496 100268622 100268890 100269103 100269151 100269320 100269607 100269777 100269817 100270029 100270283 100270481 100263198 100263538 100263582 100263699 100264439 100265430 100265697 100266015 100266153 100266264 100266520 100266683 100266771 100267091 100267308 100267438 100267570 100267845 100267986 100268190 100268329 100268497 100268623 100268891 100269104 100269152 100269321 100269608 100269778 100269889 100270030 100270284 100270482 100263199 100263539 100263583 100263701 100264440 100265432 100265698 100266016 100266154 100266265 100266521 100266685 100266872 100267092 100267309 100267439 100267571 100267846 100267987 100268191 100268330 100268498 100268624 100268892 100269105 100269153 100269322 100269609 100269779 100269890 100270031 100270285 100270483 100263280 100263540 100263584 100263703 100264441 100265433 100265699 100266017 100266155 100266266 100266522 100266686 100266873 100267093 100267310 100267440 100267572 100267847 100267988 100268192 100268331 100268499 100268625 100268893 100269106 100269154 100269323 100269610 100269780 100269891 100270032 100270286 100270484 100270485 100270774 100270968 100271029 100271321 100271586 100271738 201035373 201035458 201035506 201035563 100270486 100270775 100270969 100271030 100271322 100271587 100271739 201035374 201035459 201035507 201035564 100270487 100270776 100270970 100271031 100271323 100271588 100271800 201035395 201035460 201035508 201035565 100270488 100270777 100270971 100271032 100271324 100271590 100271801 201035396 201035461 201035509 201035566 100270660 100270778 100270972 100271033 100271325 100271591 100271802 201035397 201035462 201035510 201035567 100270661 100270779 100270973 100271034 100271326 100271592 100271803 201035398 201035463 201035511 201035568 100270662 100270780 100270974 100271035 100271327 100271593 100271804 201035399 201035464 201035512 201035569 100270663 100270781 100270975 100271036 100271328 100271594 100271805 201035423 201035465 201035513 201035570 100270664 100270782 100270976 100271037 100271329 100271595 100271806 201035424 201035466 201035514 201035571 100270665 100270783 100270977 100271038 100271330 100271596 100271807 201035425 201035467 201035515 201035572 100270666 100270784 100270978 100271039 100271331 100271597 100271808 201035426 201035469 201035516 201035573 100270667 100270785 100270979 100271113 100271332 100271598 100271809 201035428 201035470 201035517 201035574 100270668 100270786 100270980 100271114 100271333 100271599 100271905 201035430 201035471 201035518 201035575 100270669 100270787 100270981 100271115 100271334 100271660 100271906 201035431 201035472 201035519 201035576 100270670 100270788 100270982 100271116 100271335 100271661 100271907 201035432 201035473 201035520 201035577 100270671 100270789 100270983 100271117 100271336 100271662 100271908 201035433 201035474 201035521 201035578
US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam
2851