Class II
Terminated
OT1000 Series Orthopedic Surgical Tables
Recall: Z-1149-2021
·
Reported March 3, 2021
Enforcement
- Recall Number
- Z-1149-2021
- Event ID
- 87230
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Steris Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 3, 2021
- Initiation Date
- January 21, 2021
- Classification Date
- February 24, 2021
- Termination Date
- November 10, 2021
- Address
- 5960 Heisley Rd, N/A, Mentor, OH, 44060-1834, United States
Description
OT1000 Series Orthopedic Surgical Tables
Reason
On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Code Info
Model No.: OT1000, OT1100, OT1200
Distribution
Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.
Quantity
272 tables total (Domestic: 227 tables, Foreign 45 tables)