FDA Enforcement Class I Ongoing

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-00-0800-55

Recall: Z-1145-2023 · Reported March 15, 2023

Enforcement

Recall Number
Z-1145-2023
Event ID
91602
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Datascope Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 15, 2023
Initiation Date
February 7, 2023
Classification Date
March 9, 2023
Address
1300 Macarthur Blvd, Mahwah, NJ, 07430-2052, United States

Description

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-00-0800-55

Reason

An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.

Code Info

All serial numbers. Model 0998-00-0800-31, UDI-DI 10607567109053; Model 0998-00-0800-32, UDI-DI 10607567111117; Model 0998-00-0800-33, UDI-DI 10607567109008; Model 0998-00-0800-34, UDI-DI 10607567111940; Model 0998-00-0800-35, UDI-DI 10607567109107; Model 0998-00-0800-45, UDI-DI 10607567108421; Model 0998-00-0800-52, UDI-DI 10607567108438; Model 0998-00-0800-53, UDI-DI 10607567108391; Model 0998-00-0800-55, UDI-DI 10607567108414; Model 0998-00-0800-65, UDI-DI 10607567113432

Distribution

Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.

Quantity

4062 Total