Allura Xper FD20; Catalog numbers: (1) 722006, (2) 722012, (3) 722028.
Enforcement
- Recall Number
- Z-1144-2025
- Event ID
- 96132
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 26, 2025
- Initiation Date
- January 13, 2025
- Classification Date
- February 20, 2025
- Address
- Veenpluis 4-6, N/A, Best, N/A, N/A, Netherlands
Description
Allura Xper FD20; Catalog numbers: (1) 722006, (2) 722012, (3) 722028.
Possibility of the patient falling from the table related to the mattress used on the patient table.
Catalog numbers: (1) 722006, (2) 722012, (3) 722028; UDI-DI: (1) No UDI, (2) 00884838059054, (3) 00884838054202; Serial Numbers: (1) 2013, 2155, 1473, 2110, 1364, 2268, 172, 2324, 1637, 1881, 2229, 2533, 2158, 1774, 468, 813, 1477, 2051, 2462, 1352, 2562, 771, 1734, 1589, 1445, 1547, 929, 1414, 731, 2131, 635249, 2569, 1461, 914, 2097, 1890, 2281, 1896, 1762, 1764, 159, 774, 2198, 2009, 1629, 810, 1729, 2454, 1209, 1646, 1924, 1339, 2301, 1927, 1928, 1706, 1705, 2571, 2524, 1582, 972, 878, 2200, 1575, 707, 1761, 861, 2537, 722, 72246F, 1674122, 1730, 1936, 1490, 2045, 2536, 1338, 1659, 2610, 920, 1662, 1346, 1587, 1977, 2140, 664, 706, 1843, 1594, 1875, CV.381, 886, 2523, 2267, 2823014, 1583, 1585, 1520, 2126, 1856, 1941, 288, 2390, 1425, 1496, 862, 896, 2608, 1393, 2204, 2366, 2449, 2456, 2269, 2425, 2066, 1744, 1747, 1748, 1325, 1812, 2606, 641, 1728, 559, 1443, 2530, CV.271, 283, 1755, 1120, 889, 518, 1007, 1104, 1914, 1752, 2565, 848, 904, 1754, 1375, 1696, 843, 2021, 2195, 1723, 767, 2585, 1523, 1733, 1978, 2102, 1085, 18268, 1588, 806, 782, 1177, 587, 161, 1391, 1956, 638, 1387, 2027, 1601, 2561, 2256, 2070, 2002, 1627, 893, 2347, 2548, 2587, 2427, 1865, 2152, 1361, 91, 2430, 2028, 2312, 1711, 2003, 1370, 1484, 1564, 1534, 1987, 2054, 2283, 1888, 1854, 1407, 1641, 852, 853, 2062, 1366, 1604, 2568, 1169, 1114, 1115, 1116, 1127, 01H041985, 723, 2106, 2111, 2112, 1536, 1732, 1870, 1420, 1818, 1775, 1341, 1827, 849, 1550, 1359, 913, 674, 1720, 1476, 1766, 1446, 1908, 2326, 828, 2525, 1289, 1306, 830, 1781, 2296, 1972, 1932, 1935, 1273, 1502, 2201, 804, 1727, 1900, 2088, 1480, 1168, 1631, 1196, 1831, 2007, 2187, 797, 1328, 1922, 1384, 2443, 1844, 1848, 1611, 683, 2143, 1323, 778, 1081, 1466, 2336, 327, 1469, 1830, 38925, 1832, 1600, 775, 1654, 1707, 2459, 2297, 1639, 1153, 1167, 1615, 1833, 992, 1612, 1939, 2184, 1807, 1872, 2042, 2489, 2181, 1448, 1868, 1955, 1740, CV.514, 888, 2380, 2272, S4F621389A000001, 1050, 1511, 550, 2147, 1942, 2382, 1274, 1276, 1500, 1845, 1441, 1438, 658, 1518, 2250, 1780, 1326, 1259, 1913, 91A, 1731, 1000, 1653, 2203, 680, 1221, 2199, 1609, 599, 1921, 2135, 1886, 2378, 663, 1889, 2183, 1433, 80320, 2025, 2576, 1911, 1648, 1174, 1507, 1229, 1434, 2323, 1318, 1722, 1560, 1093, 1750, 2564, 1205, 781, 722008-358, 1236, 1710, 2058, 2161, 2612, 1603, CV.740, 1621, 1640, 1670, 2159, 1578, 1906, 602A, 1210, 1383, 2412, 1374, 2274, 2286, 1983, 1372, 1373, 1838, 1270, 1591, 2318, 1022, 1246, 1859, 2371, 1230, 1580, 1745, 1464, 1724, 2232, 2194, 2046, 2535, 2231, 1403, 2117, 2389, 2043, 958, 1861, 1224, 1113, 1255, 2173, 1402, 1712, 666, 2521, 589, 2236, 811, 1200, 1505, 2416, 1714, 1336, 2349, 1650, 1787, 2214, 1538, 1919, 854, 1899, 1204, 1106, 1961, 2150, 2255, 1506, 1651, 2392, 1808, 1815, 1463, 1814, 1678, 2039, 1840, 2144, 1991, 1819, 1880, 1985, 1307, 867, 1988, 1529, 2041, 2104, 1515, 1892, 2314, 2063, 1003, 2101, 2488, 2438, 1031, 1992, 1715, 2095, 1343, 2093, 1347, 1784, 955, 2094, 1284, 2115, 1530, 1759, 1344, 2245, 1280, 2067, 946, 1316, 1817, 2162, 2348, 1130, 1232, 2061, 833, 884, 2288, 2292, 1231, 1537, 1295, 1967, 2399, 873, 856, 2423, 1816, 1554, 1916, 1796, 812, 2237, 1763, 2307, 2064, 1561, 2244, 2361, 2004, 1793, 1617, 952, 837, 1285, 802, 1746, 1947, 2325, 1998, 1348, 2360, 2293, 1233, 2291, 2022, 1893, 2434, 1968, 887, 1849, 2137, 1065, 2369, 2109, 2192, 1004, 2393, 2305, 1948, 1777, 1513, 820, 951, 925, 2080, 2145, 2023, 971, 1997, 1647, 2428, 1704, 492, 2247, 2240, 1160, 1999, 1424, 2026, 1468, 1198, 1253, 2015, 2555, 2368, 2553, 2538, 2034, 2550, 1427, 533, 2551, 1337, 2584, 922, 1794, 2090, 164, 2514, 880, 1791, 2419, 1602, 1824, 1397, 1980, 1823, 1668, 2526, 2426, 2552, 2465, 2541, 2529, 491, 585, 86, 1632, 63, 2171, 973, 1590, 1598, 2285, 1795, 2049, 1949, 1595, 2079, 1568, 1392, 2206, 2343, 2411, 2124, 1349, 1454, 1455, 1456, 1751, 2031, 963, 1658, 885, 2153, 2310, 931, 2010, 1749, 2275, 1225, 1549, 2353, 940, 1330, 1952, 2379, 2598, 640, 936, 2136, 2367, 1545, 2209, 2600, 890, 2493, 2375, 602, 799, 1540, 1556, 1345, 2096, 866, 1969, 1542, 1933, 2259, 296, 1876, 2217, 2085, 1021, 1452, 2103, 1552, 1460, 1917, 1667, 2038, 1222, 2520, 1180, 2475, 2176, 1686, 1687, 2129, 2342, 8, 1075, 2086, 689, 1509, 1708, 2175, 2179, 808, 1401, 2163, 1080, 1149, 1533, NL94, 1219, 1220, 724, 1965, 2154, 2230, 1773, 1858, 2182, 2334, 1618, 1990, 1811, 1129, 2132, 1405, 1771, 1864, 1281, 1296, 2165, 1250, 1757, 1488, 2166, 2290, 1056, 520, 1738, 617, 1690, 717, 609, 1925, 819, 2040, 2069, 2211, 1478, 687, 2531, 1079, 1524, 1851, 1753, 2528, 1492, 578, 2532, 989, 1517, 2527, 1440, 2243, 2580, 2517, 1423, 636, 1557, 1408, 990, 1312, 2225, 618, 794, 1741, 2391, 1758, 2373, 2014, 2114, 1802, 1525, 2219, 2228, 1801, 2248, 2279, 1871 ****Serial Numbers Added 5/2/2025**** 296, 1674122, 38925, 2576, 2122, 1972, 1812, 1775, 1722, 1705, 1461, 1434, 1325, 1210, 1205, 1196, 1167, 1153, 896, 813; (2) 2353, 919, 1581, 1117, 1673, 1712, 585, 2876, 517, 1545, 70, 1715, 504, 757, 935, 1356, 1250, 1613, 2006, 1440, 1550, 1733, 1734, 2201, 1922, 741, 2283, 92, 1034, 1359, 1848, 950, 2943, 865, 1908, 1166, 1113, 1659, 494, 335, 1804, 1473, 100, 1408, 1401, 1588, 576, 595, 590, 2302, 2508, 2511, 1279, 2509, 452, 2151, 1724, 1609, 2896, 2134, 481, 348, 543, 893, 1187, 1455, 2085, 2202, 423, 718, 1282, 325, 2723, 2697, 2721, 2722, 2726, 796, 2890, 525, 2307, 1798, 1656, 399, 990, 2693, 1806, 1871, 2506, 2518, 243, 951, 1652, 2883, 894, 1597, 596, 1054, 1189, 1099, 823, 1037, 566, 72, 1606, 924, 2310, 2713, 1610, 686, 2933, 2514, 1108, 2829, 1294, 568, 500, 459, 469, 2886, 2760, 1077, 440, 1422, 1537, 2875, 2888, 1035, 285, 84, 1317, 1649, 1895, 1900, 1723, 878, 2221, 2496, 2311, 2222, 1515, 2828, 1803, 2484, 1731, 1017, 1020, 380, 1811, 527, 2171, 1497, 782, 418, 1053, 1558, 1336, 601, 1014, 684, 1265, 331, 389, 732, 751, 1324, 2276, 2086, 1285, 2219, 2687, 298, 814, 815, 1120, 278, 297, 1713, 1506, 1098, 1852, 1307, 1858, 377, 758, 1598, 2912, 172, 2054, 856, 1690, 1407, 480, 690, 579, 1327, 695, 1207, 1036, 2740, 733, 1091, 1254, 797, 213, 735, 1280, 2138, 828, 2821, 1118, 740, 1249, 542, 1033, 2265, 2935, 490, 470, 1075, 2203, 2774, 1321, 1396, 606, 444, 66, 2092, 597, 1269, 67, 244, 1002, 1022, 1023, 2264, 2234, 1805, 1192, 1823, 483, 484, 615, 2150, 2899, 892, 1478, 1262, 1688, 1961, 1732, 1062, 649, 650, 767, 2657, 1104, 1641, 1277, 2609, 1975, 276, 316, 959, 1335, 1461, 159, 489, 766, 639, 669, 671, 1583, 451, 625, 2692, 651, 1047, 107, 1447, 99, 670, 676, 124, 1616, 1660, 1083, 1807, 1924, 511, 1714, 122, 101, 75, 1406, 352, 492, 472, 642, 643, 1270, 577, 578, 1306, 855, 1456, 2269, 71, 1078, 768, 2285, 1793, 1797, 1216, 948, 387, 498, 1570, 2521, 1662, 104, 1868, 2313, 1743, 369, 2516, 1239, 1561, 464, 2286, 2293, 2901, 1960, 616, 1508, 1084, 2289, 1038, 1058, 318, 655, 1796, 1026, 2629, 1337, 685, 1560, 1441, 2926, 1854, 2288, 473, 2068, 624, 531, 1586, 1617, 1904, 1354, 1918, 1906, 160, 383, 560, 68, 2755, 2614, 114, 1948, 1195, 518, 888, 1329, 1992, 323, 792, 1003, 1219, 2898, 89, 115, 2003, 2004, 2835, 2843, 135, 174, 1832, 2133, 2166, 562, 1941, 760, 859, 1593, 1608, 1947, 1950, 731, 520, 848, 2292, 2656, 1946, 274, 1063, 611, 1587, 1226, 613, 1271, 1842, 205, 2736, 1799, 1665, 187, 1513, 2889, 2868, 2869, 2015, 2207, 2840, 2841, 2842, 2849, 2309, 319, 2161, 1116, 691, 1993, 208, 943, 2873, 150, 1281, 1945, 1919, 110, 337, 1735, 1326, 788, 488, 2002, 132, 1218, 1519, 1533, 336, 694, 1334, 2013, 1405, 1141, 923, 925, 533, 406, 2741, 1293, 513, 2712, 1446, 1625, 1675, 1683, 1148, 1181, 207, 442, 1305, 790, 499, 1726, 85, 476, 1657, 1531, 1630, 2610, 851, 1827, 277, 1923, 1516, 2677, 404, 453, 321, 334, 197, 1869, 2215, 2857, 574, 1611, 2874, 1559, 2924, 1686, 2855, 2856, 1635, 858, 1224, 1345, 86, 2057, 1430, 2696, 1255, 1599, 302, 610, 1209, 920, 666, 1989, 2844, 2014, 2022, 1126, 1549, 1943, 1188, 1073, 1404, 1664, 103, 1272, 2063, 2319, 594, 1467, 1596, 2607, 326, 926, 675, 756, 1027, 424, 1894, 2145, 2146, 2148, 234, 2065, 1527, 1477, 2227, 2900, 1177, 1645, 2522, 2735, 93, 1167, 1595, 2469, 88, 1672, 1210, 1512, 1101, 787, 1399, 123, 126, 136, 2658, 1119, 2655, 952, 953, 954, 1005, 1661, 2156, 320, 570, 2707, 1622, 1115, 1316, 1212, 975, 976, 1032, 986, 971, 1594, 233, 239, 640, 446, 2929, 69, 1301, 2254, 2323, 430, 1542, 76, 1607, 2454, 2149, 1211, 1004, 2046, 974, 2305, 1044, 2881, 505, 2937, 310, 192, 314, 852, 2930, 224, 1016, 2645, 2653, 1097, 1532, 1964, 2135, 1357, 589, 1092, 248, 846, 963, 964, 1658, 412, 1344, 1902, 328, 781, 1909, 1132, 1361, 1504, 1433, 229, 142, 962, 1260, 2108, 1094, 837, 333, 1681, 2282, 1815, 922, 1333, 774, 1492, 2047, 1655, 1855, 1817, 309, 1052, 1994, 1428, 1562, 627, 2005, 930, 1956, 967, 2168, 910, 1168, 167, 931, 2098, 1421, 1802, 632, 137, 957, 863, 289, 1040, 409, 1072, 1332, 2042, 57, 995, 993, 516, 2350, 759, 350, 1460, 1872, 1971, 550, 2107, 914, 2352, 1156, 180, 2314, 1464, 347, 2335, 139, 795, 1865, 628, 583, 809, 194, 170, 2406, 861, 351, 1291, 2244, 2169, 482, 1363, 1457, 288, 553, 462, 1980, 203, 1580, 395, 1423, 548, 2455, 1887, 2456, 1157, 1485, 2744, 876, 1592, 582, 2346, 2747, 2074, 2072, 1256, 2743, 2714, 166, 746, 2631, 2418, 436, 2386, 1214, 2749, 2674, 2739, 2196, 619, 199, 1833, 2587, 2032, 630, 2584, 1825, 1839, 1591, 261, 291, 2578, 2462, 502, 60, 2547, 2548, 381, 2333, 1986, 2599, 98, 1196, 1495, 2485, 2139, 2490, 477, 2200, 779, 2734, 56, 2331, 2385, 296, 2756, 770, 507, 727, 799, 279, 992, 2813, 1413, 58, 128, 392, 541, 1480, 2553, 2545, 978, 2356, 2784, 1297, 2026, 652, 1150, 59, 1349, 911, 152, 2450, 2585, 215, 871, 2640, 2790, 2830, 1678, 2683, 2729, 2029, 2413, 2451, 2622, 1228, 2071, 2076, 2127, 1671, 2737, 1081, 1298, 1028, 2771, 2056, 1366, 1708, 2775, 275, 830, 2479, 1978, 2579, 1244, 829, 2659, 2363, 891, 2102, 2300, 663, 2675, 2474, 2757, 839, 91, 2298, 1151, 1152, 2565, 1626, 2403, 1882, 569, 688, 2016, 2780, 2181, 2296, 544, 1481, 338, 2797, 1390, 2473, 419, 2504, 2197, 1436, 2336, 728, 785, 355, 1698, 165, 793, 794, 1556, 1654, 1801, 2539, 1736, 2601, 2702, 2604, 2758, 1967, 2096, 622, 210, 2678, 2365, 2837, 2366, 2291, 2299, 2665, 2501, 1878, 2811, 1669, 2634, 648, 1471, 2704, 2765, 737, 1879, 581, 1397, 2121, 1937, 2661, 2388, 382, 2180, 1342, 647, 1818, 2673, 2786, 2651, 867, 1816, 1288, 1489, 2725, 1178, 2544, 1874, 2911, 2349, 2724, 2175, 479, 2537, 2176, 1175, 1237, 2691, 1381, 2538, 2274, 2715, 398, 2523, 2491, 2763, 2785, 2459, 2716, 1965, 2160, 941, 211, 1147, 1303, 803, 1699, 1130, 798, 441, 1089, 2272, 2429, 2788, 1463, 1479, 2685, 534, 1138, 1203, 1828, 2401, 752, 2824, 2738, 870, 2344, 1536, 2249, 2612, 2532, 2410, 772, 2781, 2120, 1021, 1382, 2441, 2326, 2345, 1999, 2405, 2766, 1309, 2438, 1739, 1572, 2173, 2174, 1836, 2558, 2343, 2826, 2831, 680, 980, 2058, 2568, 1142, 2770, 2549, 1263, 2695, 2583, 730, 2031, 1976, 2118, 2810, 1079, 1530, 1875, 593, 2649, 2394, 1808, 2527, 1412, 506, 1548, 2384, 280, 2718, 2153, 748, 188, 2140, 2576, 2476, 2666, 1567, 2411, 2431, 2433, 2834, 1064, 1917, 1287, 1795, 1137, 635, 2358, 2798, 509, 2617, 2287, 2728, 2577, 1484, 667, 2795, 2524, 2719, 2808, 2708, 2443, 2348, 2442, 2641, 1411, 2444, 1127, 2330, 659, 2624, 528, 2626, 1711, 2128, 2581, 2778, 2792, 1056, 1134, 1862, 2470, 1348, 2481, 2711, 2777, 2806, 806, 2396, 1241, 223, 407, 1898, 2460, 2179, 2467, 1061, 443, 405, 1143, 665, 2493, 2494, 1169, 1223, 629, 2773, 763, 260, 1535, 2393, 2615, 125, 432, 966, 2380, 987, 1631, 2638, 2637, 2379, 2557, 2525, 2392, 2745, 2551, 2720, 1233, 2290, 2776, 2632, 2030, 2759, 2825, 2482, 1234, 1088, 1927, 2117, 2017, 106, 2648, 2466, 1966, 2540, 2573, 2398, 2409, 2507, 599, 1705, 1721, 2427, 2566, 2812, 753, 1624, 2387, 478, 184, 2424, 266, 2799, 1462, 546, 1289, 2764, 810, 771, 2752, 2690, 2853, 2550, 2122, 1296, 755, 2543, 2647, 1834, 2430, 2526, 2208, 2701, 2761, 287, 2854, 743, 2633, 2475, 1347, 1977, 1703, 2461, 1568, 2779, 2681, 693, 902, 2613, 2555, 2662, 2845, 2270, 917, 2371, 2689, 2575, 1958, 2342, 2832, 2768, 1131, 2580, 2155, 2646, 1174, 1375, 982, 1268, 2688, 2101, 662, 556, 2487, 1885, 2650, 2341, 2449, 1997, 2530, 2556, 2028, 979, 573, 955, 2847, 2033, 2434, 2338, 2389, 2477, 679, 2528, 376, 269, 2750, 1926, 332, 916, 181, 2671, 2297, 1668, 1434, 1623, 2838, 660, 1076, 414, 2007, 2751, 936, 2390, 2453, 2531, 2351, 2546, 2376, 1074, 2552, 1086, 1059, 2616, 1814, 2301, 2559, 2589, 2188, 588, 1360, 2087, 2801, 2767, 895, 2643, 2367, 2802, 1377, 272, 216, 1860, 2574, 2372, 2373, 587, 1614, 2562, 2399, 661, 2567, 311, 554, 219, 2630, 2608, 1830, 1984, 2205, 1242, 390, 1201, 813, 1740, 2471, 2250, 1819, 2327, 1071, 2542, 540, 127, 1128, 2412, 2746, 903, 2836, 2884, 1701, 572, 2407, 2489, 1861, 1695, 2668, 726, 944, 2391, 2823, 2103, 2437, 2698, 391, 2378, 817, 524, 2782, 2623, 415, 2095, 2870, 2762, 2535, 608, 2667, 117, 2082, 2800, 2619, 116, 1144, 1145, 631, 1122, 374, 738, 2464, 2593, 2732, 2421, 227, 2252, 1222, 112, 2124, 2488, 185, 2404, 218, 2375, 2182, 400, 2340, 1202, 1693, 2080, 869, 2502, 2228, 2229, 1415, 764, 2422, 2478, 742, 2325, 2748, 2067, 1385, 2364, 1135, 1881, 2621, 2093, 2034, 539, 1835, 805, 1435, 214, 360, 2594, 1450, 2334, 555, 2536, 2492, 1284, 1266, 2094, 1364, 1093, 2008, 2793, 773, 2730, 1998, 2112, 2818, 2804, 1483, 2694, 868, 2733, 2541, 2851, 2024, 2414, 2415, 2439, 2440, 61, 1232, 1737, 2618, 1702, 1741, 2827, 2682, 2679, 410, 2789, 2554, 2198, 2278, 833, 901, 1267, 271, 2676, 2591, 388, 2123, 1653, 2852, 2177, 1367, 2432, 2195, 2611, 1388, 2803, 2495, 1391, 1972, 2794, 2590, 2636, 2073, 1996, 2382, 2791, 1261, 175, 176, 557, 222, 840, 437, 1082, 426, 1633, 672, 1379, 872, 692, 1312, 2077, 1820, 1973, 144, 1821, 1822, 972, 161, 1453, 2887, 2880, 1837, 73, 74, 1696, 118, 119, 2045, 2919, 2043, 678, 2891, 780, 155, 140, 1292, 1290, 2703, 2517, 246, 762, 34, 1939, 2684, 25, 2520, 154, 1585, 2877, 2206, 177, 835, 598, 1245, 1930, 1897, 2144, 2304, 1246, 1494, 2664, 880, 1158, 1951, 897, 2137, 523, 2519, 1663, 2921, 1153, 2916, 609, 2000, 1400, 2914, 403, 26, 1013, 1451, 2934, 148, 385, 1748, 1845, 1425, 1988, 2312, 111, 777, 749, 1720, 1546, 393, 592, 1870, 2303, 510, 80, 1838, 2256, 2172, 623, 1544, 2316, 2220, 2639, 1985, 1049, 687, 1915, 1358, 674, 1829, 1710, 1987, 2294, 918, 602, 428, 1172, 1322, 745, 1891, 1979, 808, 928, 1719, 465, 1024, 988, 824, 1582, 807, 1048, 529, 617, 857, 1346, 834, 985, 983, 2754, 1259, 354, 1350, 163, 138, 2627, 1620, 262, 1121, 421, 1136, 2927, 2918, 1180, 2011, 1452, 1534, 1813, 143, 1176, 1520, 273, 1949, 1955, 1867, 105, 1639, 1850, 1541, 1901, 536, 2266, 2497, 2280, 940, 1015, 657, 1198, 1634, 1742, 1899, 1338, 2498, 1454, 1553, 2663, 2625, 178, 1840, 212, 2066, 2458, 2499, 1619, 1812, 1055, 1251, 1418, 956, 249, 1417, 1368, 2090, 887, 2100, 1543, 1744, 933, 2772, 1676, 1230, 845, 2110, 2141, 2505, 292, 1547, 1962, 1124, 113, 2922, 2059, 1149, 1319, 2036, 468, 236, 545, 1227, 1931, 1060, 514, 1651, 491, 932, 1554, 1200, 1647, 1341, 1932, 294, 1220, 526, 1935, 769, 408, 1847, 2089, 873, 874, 173, 994, 1046, 1745, 1700, 2084, 1313, 2362, 342, 2359, 1355, 191, 537, 341, 1648, 2360, 2361, 786, 744, 1925, 373, 2097, 1171, 1551, 1794, 2091, 471, 1080, 1615, 149, 1, 618, 1439, 312, 1844, 1911, 1697, 353, 2114, 1283, 497, 1448, 1206, 267, 1123, 1974, 2253, 1330, 425, 890, 1050, 1273, 621, 1888, 818, 1907, 2882, 1374, 162, 1747, 1264, 1517, 1318, 1085, 2078, 946, 945, 1565, 226, 179, 1372, 1125, 1371, 1557, 1186, 645, 1864, 1449, 1351, 1031, 1574, 1640, 1637, 1575, 1576, 1573, 1493, 1248, 1577, 1667, 646, 1579, 1403, 2069, 1914, 329, 1173, 2652, 968, 1684, 2915, 1398, 2602, 1569, 2897, 2513, 433, 343, 2917, 2628, 413, 2879, 2913, 2706, 2606, 429, 304, 305, 306, 307, 308, 2850, 327, 1159, 2010, 1490, 1670, 921, 2347, 882, 1051, 567, 2941, 422, 1304, 1853, 1057, 1295, 2055, 463, 850, 1090, 1968, 1129, 1841, 603, 1959, 604, 454, 460, 1468, 812, 2053, 1555, 1628, 847, 1666, 1896, 1921, 1320, 1969, 2130, 183, 375, 1528, 1503, 1982, 240, 1706, 969, 411, 634, 295, 241, 242, 157, 156, 804, 658, 1068, 1432 ****Serial Numbers Added 5/2/2025**** 92, 315, 318, 472, 596, 642, 790, 951, 1054, 1226, 1238, 1272, 1286, 1401, 1473, 1588, 1606, 1659, 1692, 2002, 2951, 73, 74, 137, 149, 199, 226, 347; (3) 1811, 819, 1598, 316, 1946, 945, 541, 565, 647, 674, 864, 1041, 710, 880, 946, 949, 961, 447, 236, 237, 238, 356, 696, 1188, 1189, 1098, 793, 618, 313, 329, 274, 1214, 1846, 1978, 1649, 2192, 904, 242, 284, 1397, 1957, 167, 207, 208, 526, 118, 1820, 1756, 1677, 991, 479, 2348, 2641, 431, 2020, 2760, 601, 869, 377, 1454, 1629, 1984, 275, 2933, 52, 66, 1340, 539, 1, 272, 1103, 1311, 494, 80, 1136, 1639, 282, 1408, 1232, 990, 262, 1275, 1956, 406, 548, 1017, 586, 530, 136, 137, 1441, 1130, 940, 1657, 938, 1656, 1736, 1076, 929, 251, 731, 989, 1834, 981, 350, 125, 368, 971, 499, 211, 261, 1424, 2647, 1856, 1261, 924, 934, 1854, 721, 1142, 1002, 1651, 1385, 774, 2174, 1930, 2846, 1463, 1508, 892, 779, 1111, 518, 519, 590, 1058, 452, 453, 811, 1370, 543, 680, 1074, 668, 1191, 1212, 11, 318, 1211, 343, 424, 435, 437, 340, 481, 806, 1837, 805, 468, 906, 317, 1774, 2502, 1132, 1081, 982, 592, 655, 143, 1405, 1482, 140, 1413, 1448, 464, 2103, 1213, 139, 552, 303, 476, 419, 521, 522, 531, 1201, 1014, 861, 580, 1327, 256, 917, 1266, 1472, 2083, 2007, 2008, 1652, 1129, 818, 1394, 1396, 1646, 254, 326, 122, 1116, 1824, 324, 1716, 130, 2295, 1288, 2906, 877, 896, 987, 995, 1755, 928, 267, 759, 866, 1310, 895, 1407, 1187, 128, 273, 1171, 888, 1264, 1222, 2843, 336, 375, 887, 449, 516, 533, 408, 373, 154, 956, 496, 1549, 376, 744, 1034, 1054, 473, 808, 630, 785, 379, 553, 110, 2261, 983, 627, 903, 1099, 1909, 654, 1481, 784, 2061, 422, 525, 363, 1690, 1931, 260, 1233, 2206, 1122, 2275, 2310, 2506, 2200, 900, 339, 781, 1850, 809, 1331, 365, 571, 203, 204, 1538, 678, 528, 560, 310, 743, 445, 161, 189, 2635, 292, 1466, 1200, 1374, 720, 1836, 173, 620, 1085, 361, 281, 475, 1094, 669, 1049, 456, 315, 908, 672, 1350, 1723, 127, 132, 141, 145, 180, 1001, 463, 1484, 1835, 1244, 1022, 1004, 1456, 538, 2173, 210, 564, 558, 407, 1050, 1751, 2284, 697, 364, 520, 2070, 1000, 582, 1349, 1351, 879, 930, 932, 882, 129, 218, 451, 485, 1393, 176, 257, 1920, 962, 1809, 556, 1353, 998, 691, 215, 279, 459, 976, 1011, 1267, 2196, 889, 729, 1149, 1279, 1095, 1870, 2253, 347, 909, 1556, 1498, 1650, 952, 160, 540, 1569, 306, 234, 213, 692, 925, 977, 1202, 1566, 222, 596, 994, 1269, 131, 133, 138, 1855, 224, 48, 1783, 2834, 490, 968, 2645, 67, 1100, 1570, 2249, 1352, 1869, 638, 1447, 631, 637, 2151, 2601, 570, 529, 493, 534, 1326, 2710, 491, 867, 891, 2110, 2745, 31, 206, 751, 1890, 2293, 561, 205, 305, 488, 1822, 1880, 209, 825, 1461, 2707, 2709, 579, 1634, 1077, 1875, 498, 462, 677, 757, 1948, 1043, 142, 280, 159, 1373, 1949, 169, 2922, 535, 2252, 386, 158, 922, 184, 185, 1154, 489, 1826, 466, 671, 978, 1499, 1708, 1710, 217, 194, 195, 197, 216, 227, 402, 1483, 559, 967, 973, 1068, 1069, 1032, 454, 918, 1705, 821, 914, 698, 1387, 854, 1051, 1988, 675, 741, 768, 910, 666, 1889, 1983, 1568, 1845, 1980, 547, 550, 416, 1168, 394, 428, 636, 901, 1446, 1828, 1368, 426, 1035, 1995, 684, 1333, 1457, 1514, 190, 840, 842, 625, 1630, 166, 346, 2615, 1156, 632, 1633, 71, 1567, 2407, 2422, 1395, 1428, 382, 2071, 121, 549, 874, 229, 228, 333, 372, 151, 695, 371, 911, 1619, 1030, 1155, 717, 265, 1343, 1439, 1600, 758, 942, 1039, 742, 878, 763, 231, 1572, 663, 1056, 905, 1007, 64, 628, 1172, 1532, 822, 430, 200, 212, 63, 1254, 1354, 788, 1292, 1008, 1046, 944, 144, 607, 1301, 1006, 1955, 870, 1199, 505, 589, 1059, 1060, 2924, 770, 59, 997, 748, 247, 524, 460, 114, 87, 1304, 1042, 1898, 2355, 2357, 2356, 728, 1088, 951, 440, 536, 264, 30, 298, 299, 300, 1816, 1865, 1692, 1819, 1817, 1818, 1925, 2104, 1146, 1554, 993, 2334, 396, 1953, 405, 1460, 1775, 2169, 639, 467, 546, 1255, 1814, 1830, 1378, 2372, 2373, 436, 827, 1390, 855, 1645, 823, 2216, 2408, 1336, 2223, 2464, 2208, 2888, 1878, 2715, 2296, 1961, 2201, 1478, 2937, 2351, 1894, 1537, 2648, 1138, 2793, 2773, 2353, 2044, 1553, 1971, 2014, 2820, 1402, 2406, 2856, 1696, 2317, 2416, 1518, 1511, 2254, 2899, 2405, 2243, 2251, 2301, 835, 1540, 1167, 2215, 1761, 1724, 2675, 2362, 2625, 2175, 2956, 2437, 1315, 948, 2619, 2639, 2497, 1923, 2126, 1654, 1218, 1805, 1739, 1746, 2935, 2488, 2860, 1864, 2850, 1991, 1992, 2279, 2626, 2229, 2094, 1762, 2016, 2398, 2659, 2676, 2383, 1643, 2230, 2232, 2157, 2182, 2451, 2508, 1091, 2548, 2787, 2471, 2233, 2036, 1223, 2921, 2341, 2018, 2515, 2087, 2950, 1906, 2630, 2388, 2486, 2824, 2944, 2740, 2644, 2132, 2678, 2778, 2702, 2327, 1664, 2394, 2448, 1576, 1954, 2202, 1599, 2307, 1793, 2198, 2792, 2289, 2000, 1422, 2361, 1942, 2207, 2430, 2264, 1839, 1489, 2128, 1994, 2461, 1623, 1624, 2822, 2325, 2823, 2463, 760, 2389, 2220, 1564, 2359, 2901, 1797, 2332, 2382, 1849, 2042, 2475, 2272, 2941, 2012, 1742, 2501, 2469, 2184, 1888, 2440, 2484, 2302, 1410, 2303, 2712, 1887, 1185, 2013, 1341, 2825, 1901, 2872, 1941, 1867, 2429, 2218, 2840, 2160, 1744, 2505, 2333, 2663, 2603, 2434, 2159, 1057, 2379, 2255, 2794, 2031, 2705, 1969, 2248, 936, 1300, 2121, 2097, 1769, 837, 1903, 1681, 2562, 2544, 2561, 2518, 2801, 2868, 2646, 2412, 1661, 2893, 1162, 1127, 2661, 1376, 2754, 2473, 1561, 2516, 2368, 2393, 999, 1589, 1738, 2219, 1905, 2139, 2564, 2345, 2730, 1586, 2599, 1145, 1192, 2411, 1808, 1848, 1276, 2890, 2851, 2049, 2015, 1981, 2022, 2084, 2433, 2432, 2717, 2931, 2390, 1773, 2245, 1544, 1940, 1933, 2897, 1863, 2455, 1916, 2495, 1471, 2210, 2837, 2391, 2565, 2453, 2748, 1535, 1699, 2833, 1802, 2242, 1706, 2703, 2927, 2681, 2278, 2156, 1778, 2802, 2904, 2131, 1627, 2566, 2454, 1963, 2423, 1934, 2023, 1851, 1678, 2082, 2948, 2734, 1086, 1087, 1121, 1527, 2812, 1668, 2449, 2830, 2919, 1506, 2125, 2360, 2262, 1904, 2640, 2952, 1228, 1504, 1752, 2584, 2442, 2118, 2399, 1528, 2577, 2947, 2158, 2666, 2152, 2768, 2445, 2918, 2739, 2829, 2757, 2580, 2928, 2444, 1868, 2080, 1622, 2945, 1734, 1764, 2657, 2817, 1038, 2816, 2714, 1113, 2450, 1473, 2942, 2004, 2900, 1784, 2582, 2629, 2838, 1977, 2011, 2946, 2462, 2803, 2419, 2420, 1253, 2867, 1832, 2542, 2340, 2827, 1555, 2799, 2046, 2658, 2109, 2667, 2875, 2089, 1997, 2336, 1727, 2155, 1939, 1465, 2209, 1290, 2880, 2529, 2592, 2401, 2427, 2669, 2857, 2907, 1416, 1659, 487, 2624, 2769, 2321, 2810, 1573, 1503, 2591, 1683, 2524, 2689, 2682, 2509, 1317, 2798, 2557, 2595, 2096, 2672, 1680, 2543, 1730, 1250, 2001, 2041, 2538, 1615, 2716, 1324, 1929, 1490, 1936, 2064, 2651, 1502, 1647, 2482, 1987, 2704, 2869, 848, 2696, 1398, 2706, 2441, 1772, 1383, 1530, 2312, 1873, 2578, 1477, 2358, 1965, 2616, 2068, 2756, 2111, 2512, 1776, 1796, 2291, 2795, 1382, 2774, 2460, 1711, 2322, 2259, 2431, 2859, 1861, 2569, 1469, 2299, 2066, 1468, 1871, 2320, 2520, 2344, 2751, 2078, 2380, 2700, 1841, 2593, 2519, 1891, 1763, 1935, 2350, 2113, 1838, 2797, 2873, 2759, 2298, 1541, 2428, 2891, 2308, 2732, 2628, 2673, 2257, 2749, 1791, 2766, 2701, 1876, 2679, 2540, 2140, 2788, 2826, 1719, 2069, 1765, 2328, 2742, 1766, 2597, 2818, 1375, 2746, 2470, 2558, 2177, 2311, 2417, 2032, 2535, 2415, 2841, 2831, 1384, 1747, 2148, 2185, 1781, 2288, 1161, 2836, 1516, 2117, 1594, 2762, 2136, 2552, 2571, 2631, 1777, 2934, 2882, 2145, 2845, 1229, 2392, 2650, 2369, 1316, 2221, 1874, 1745, 1877, 1474, 1844, 2149, 2863, 2349, 2568, 2627, 2052, 1847, 2951, 2028, 2374, 2499, 1355, 1593, 2721, 1493, 2807, 2534, 2079, 1595, 2002, 2943, 2575, 2114, 2439, 2436, 1924, 2728, 2684, 1862, 1801, 2147, 1449, 2889, 2602, 2459, 2784, 2832, 2347, 2178, 2938, 2457, 1520, 1164, 2339, 2466, 503, 2723, 2323, 1048, 1910, 2511, 2424, 2425, 1803, 2072, 2892, 2287, 1131, 2608, 1431, 2522, 2183, 915, 1662, 1697, 1588, 2127, 2655, 2953, 1451, 2713, 1294, 2504, 2660, 2870, 1660, 2133, 2135, 2163, 2607, 1896, 1107, 2929, 1653, 2063, 2677, 2874, 2726, 2062, 2164, 2250, 2370, 1585, 2300, 2381, 2146, 2690, 2228, 2142, 2549, 2386, 2930, 2767, 2337, 2600, 2500, 2556, 1512, 2611, 2479, 1055, 2366, 2367, 1459, 1342, 2800, 1974, 1531, 2898, 2067, 2909, 1360, 2452, 1926, 2821, 2596, 2884, 2285, 2105, 1455, 2045, 1780, 1648, 2554, 2887, 2503, 2546, 1193, 2048, 2075, 2040, 1790, 2527, 2481, 2483, 1915, 2309, 2395, 2733, 1090, 2785, 2786, 1196, 2387, 2270, 1536, 2649, 2122, 2313, 2047, 1575, 2487, 2378, 1246, 2940, 2203, 2116, 1972, 2410, 1252, 2077, 2594, 2881, 2051, 2894, 1470, 2804, 2474, 2914, 2865, 510, 1753, 1357, 2536, 931, 2864, 2363, 2849, 2227, 1732, 1401, 2743, 1208, 2560, 2668, 603, 604, 2913, 1663, 2137, 1227, 2260, 2772, 2237, 2770, 1768, 2719, 2692, 2409, 2567, 876, 1767, 2204, 2205, 2671, 2741, 796, 2853, 711, 2750, 1497, 2119, 1505, 2115, 2736, 1968, 2100, 2346, 2685, 2354, 2683, 2686, 1206, 1377, 1552, 2848, 1770, 2141, 1534, 2188, 2858, 1217, 708, 2814, 2589, 1216, 2656, 2108, 2530, 2329, 2523, 1881, 2653, 1166, 1065, 2214, 1500, 2633, 1543, 1153, 1674, 1539, 1998, 2396, 2397, 2267, 2861, 1289, 1096, 2910, 2915, 1760, 815, 2421, 1670, 1671, 1682, 2521, 2081, 2574, 1976, 755, 1665, 1921, 2161, 2528, 1106, 2234, 2775, 2586, 2776, 1912, 2687, 2688, 1019, 1943, 2403, 2404, 2779, 2911, 2912, 1944, 1245, 615, 2514, 1679, 2212, 2532, 1795, 2925, 1899, 1821, 1806, 1908, 2282, 2697, 2698, 2691, 2034, 2186, 2777, 1044, 2093, 1882, 988, 2932, 2763, 2796, 1959, 2418, 2138, 2588, 2737, 2738, 2878, 2086, 2587, 2877, 2195, 2643, 2162, 2783, 2781, 2782, 943, 1464, 1885, 2130, 2476, 2477, 2839, 1902, 2708, 1571, 2271, 2902, 2903, 1962, 1295, 2021, 1427, 1804, 2165, 2095, 2828, 2193, 2576, 1725, 2263, 2213, 2326, 2003, 2091, 2120, 1728, 2854, 2545, 1990, 2835, 1429, 507, 508, 2895, 2531, 2258, 1574, 735, 736, 737, 1496, 1578, 1579, 1356, 1358, 1960, 2612, 2456, 2129, 1973, 1163, 2226, 2465, 2304, 2305, 2306, 838, 2954, 2265, 2343, 2670, 2879, 1726, 2923, 2744, 2958, 2844, 2438, 2513, 1117, 2957, 2811, 2847, 1521, 2809, 1061, 2936, 1542, 2563, 2458, 2092, 2297, 1642, 2758, 2896, 1900, 1831, 2606, 969, 1251, 2525, 1892, 2318, 2765, 1884, 2735, 2485, 2908, 1581, 1584, 2720, 2498, 2033, 2338, 1510, 2718, 2771, 2871, 1194, 1843, 2468, 2680, 2472, 1825, 1970, 2790, 2731, 2413, 2085, 2939, 2605, 1403, 2613, 2365, 1913, 2236, 2238, 2559, 1958, 907, 2634, 1415, 2197, 2319, 1717, 1886, 2780, 1409, 2553, 2331, 2006, 1026, 2144, 2268, 2722, 2517, 1240, 2269, 2693, 2623, 2694, 2695, 2747, 2225, 2808, 2480, 2866, 1158, 2166, 2154, 2330, 2187, 2526, 2352, 2886, 1325, 2027, 2755, 2764, 2283, 2852, 2949, 984, 612, 323, 853, 1082, 126, 1857, 1369, 201, 555, 702, 304, 321, 1027, 933, 415, 1487, 1488, 1932, 1577, 9, 271, 104, 38, 578, 2168, 1688, 414, 413, 13, 1911, 986, 765, 112, 824, 985, 120, 221, 512, 374, 1443, 661, 980, 164, 1347, 181, 1703, 1669, 101, 574, 747, 635, 91, 862, 514, 412, 182, 277, 1551, 330, 480, 511, 894, 232, 1332, 816, 400, 193, 780, 49, 575, 813, 709, 926, 616, 178, 233, 812, 244, 1287, 32, 1400, 35, 1979, 1104, 40, 881, 1337, 799, 163, 641, 766, 776, 1219, 1285, 297, 434, 389, 1101, 380, 624, 676, 119, 441, 1205, 1937, 105, 111, 687, 523, 2076, 576, 577, 43, 954, 192, 1907, 810, 1128, 899, 179, 1309, 54, 472, 1112, 295, 745, 123, 1207, 1307, 1308, 455, 1712, 2609, 718, 51, 777, 1517, 1735, 831, 689, 188, 730, 839, 1180, 492, 1388, 2256, 1587, 1362, 762, 1321, 851, 568, 1047, 1986, 392, 856, 1363, 2024, 1947, 1480, 1638, 1159, 1866, 168, 1097, 1025, 1438, 1184, 1596, 162, 716, 613, 857, 1580, 852, 1209, 337, 609, 1608, 1231, 1141, 249, 921, 1616, 1399, 1092, 544, 1433, 2426, 1618, 2292, 1515, 1617, 1597, 1479, 1612, 2038, 2058, 2614, 1525, 2059, 1417, 1610, 270, 1404, 1605, 1626, 1582, 970, 81, 527, 937, 1529, 767, 800, 826, 1021, 1235, 1329, 1462, 2314, 2598, 941, 2555, 1426, 2467, 2620, 2478, 2570, 2618, 1432, 679, 1045, 2621, 301, 150, 156, 2617, 919, 606, 557, 2294, 593, 594, 1548, 1361, 2101, 483, 1858, 2585, 732, 2632, 1134, 1135, 2622, 1265, 395, 1037, 98, 893, 673, 103, 622, 1996, 354, 1714, 657, 599, 693, 1257, 517, 752, 366, 1975, 633, 2199, 1550, 276, 1453, 403, 947, 898, 285, 649, 2065, 605, 427, 246, 334, 74, 653, 245, 532, 648, 199, 1210, 2153, 442, 1391, 444, 27, 1442, 1792, 484, 302, 335, 795, 773, 338, 2572, 659, 660, 863, 2019, 1418, 10, 1067, 972, 115, 640, 1271, 1066, 1458, 1895, 602, 935, 198, 290, 2400, 814, 341, 1813, 1635, 1636, 191, 1743, 1812, 2222, 875, 482, 1302, 782, 939, 617, 100, 1165, 186, 1583, 1071, 1075, 794, 432, 965, 614, 764, 501, 1695, 1694, 1704, 1693, 1313, 1312, 367, 1702, 1700, 545, 438, 1123, 1124, 832, 286, 820, 29, 802, 1258, 99, 398, 214, 665, 817, 23, 328, 608, 1064, 1320, 1089, 1070, 1621, 37, 1545, 1421, 2172, 439, 1178, 423, 235, 1632, 433, 2098, 322, 868, 1133, 1073, 694, 1346, 219, 754, 175, 202, 1221, 1105, 798, 642, 670, 727, 725, 387, 726, 722, 621, 724, 723, 2112, 1982, 2189, 1788, 2171, 974, 259, 643, 34, 1170, 2029, 314, 1297, 1799, 860, 2150, 458, 756, 1684, 410, 269, 1241, 296, 705, 703, 704, 707, 1557, 2074, 1102, 1715, 1731, 2026, 2170, 2190, 226, 1740, 847, 1918, 1625, 1685, 1118, 623, 597, 355, 1183, 2073, 291, 311, 266, 792, 1559, 1558, 551, 116, 790, 611, 1120, 79, 170, 171, 1604, 1602, 1601, 1603, 65, 753, 830, 1195, 352, 320, 2102, 690, 600, 701, 1181, 411, 1675, 153, 309, 778, 650, 662, 912, 1151, 397, 562, 353, 1152, 1952, 1148, 585, 651, 652, 248, 1815, 362, 1079, 1293, 1186, 348, 1437, 1386, 738, 369, 319, 1157, 230, 1546, 1547, 2143, 506, 2035, 1268, 77, 801, 370, 2364, 443, 183, 713, 1242, 964, 504, 187, 1914, 1789, 327, 658, 157, 883, 884, 287, 2090, 1125, 685, 834, 1893, 797, 1689, 1040, 1713, 1411, 2324, 1709, 1718, 1928, 2194, 2273, 429, 225, 588, 1879, 1389, 2099, 41, 1029, 360, 1562, 2274, 278, 1328, 1419, 955, 783, 1364, 1367, 14, 1637, 107, 124, 469, 775, 1757, 502, 2246, 1372, 388, 381, 1018, 975, 465, 803, 733, 113, 573, 844, 845, 846, 1720, 2025, 1686, 1314, 2107, 2492, 1786, 1687, 1810, 712, 1785, 2491, 1425, 714, 1613, 36, 1173, 2494, 2490, 843, 2217, 1280, 2547, 1247, 2489, 149, 2043, 1063, 2496 ****Serial Numbers Added 5/2/2025**** 305, 347, 488, 734, 909, 966, 977, 1024, 1256, 1522, 1945, 51, 59, 65, 144, 202, 246, 271, 304, 396.
Worldiwde distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.
5,029 Units (1566 US, 3463 OUS)