Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement
Enforcement
- Recall Number
- Z-1143-2024
- Event ID
- 93877
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Voco GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 28, 2024
- Initiation Date
- January 25, 2024
- Classification Date
- February 16, 2024
- Address
- Anton-Flettner-Str. 1-3, N/A, Cuxhaven, N/A, N/A, Germany
Description
Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement
Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.
a) 2115, UDI/DI +E22121151, Lot Numbers: 2346779, 2401515; b) 2117, UDI/DI +E22121171, Lot Numbers: 2347832; c) 2118, UDI/DI +E22121181, Lot Numbers: 2349536; d) 2119, UDI/DI +E22121191, Lot Numbers: 2348501; e) 2120, UDI/DI +E22121201, Lot Numbers: 2350638; f) 2122, UDI/DI +E22121221, Lot Numbers: 2349527
Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.
1021 units