FDA Enforcement Class I Terminated

BD Vacutainer¿ EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894)

Recall: Z-1141-2018 · Reported March 28, 2018

Enforcement

Recall Number
Z-1141-2018
Event ID
79223
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
March 28, 2018
Initiation Date
March 22, 2018
Classification Date
March 22, 2018
Termination Date
July 25, 2022
Address
1 Becton Dr, N/A, Franklin Lakes, NJ, 07417-1815, United States

Description

BD Vacutainer¿ EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894)

Reason

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

Code Info

All lots. Catalog Number (Shelf Life): 1. 367842 (Requested); 2. 367899 (Requested); 3. 368049 (Requested); 4. 368589 (Requested)

Distribution

Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.

Quantity

64,784,900