FDA Enforcement
Class II
Terminated
HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes.
Recall: Z-1141-2013
·
Reported April 24, 2013
Enforcement
- Recall Number
- Z-1141-2013
- Event ID
- 64723
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Horiba Instruments, Inc dba Horiba Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 24, 2013
- Initiation Date
- March 20, 2013
- Classification Date
- April 18, 2013
- Termination Date
- September 26, 2014
- Address
- 34 Bunsen, N/A, Irvine, CA, 92618-4210, United States
Description
HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes.
Reason
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.
Code Info
Model Numbers: P400ISE110EN02 and P400ISE110US02. All Serial Numbers. Versions 5.0.7 and below
Distribution
Nationwide Distribution.
Quantity
905 units