HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes.

Recall: Z-1141-2013 · Reported April 24, 2013

Enforcement

Recall Number
Z-1141-2013
Event ID
64723
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Horiba Instruments, Inc dba Horiba Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 24, 2013
Initiation Date
March 20, 2013
Classification Date
April 18, 2013
Termination Date
September 26, 2014
Address
34 Bunsen, N/A, Irvine, CA, 92618-4210, United States

Description

HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes.

Reason

HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.

Code Info

Model Numbers: P400ISE110EN02 and P400ISE110US02. All Serial Numbers. Versions 5.0.7 and below

Distribution

Nationwide Distribution.

Quantity

905 units