FDA Enforcement
Class I
Terminated
BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)
Recall: Z-1140-2018
·
Reported March 28, 2018
Enforcement
- Recall Number
- Z-1140-2018
- Event ID
- 79223
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton Dickinson & Company
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- March 28, 2018
- Initiation Date
- March 22, 2018
- Classification Date
- March 22, 2018
- Termination Date
- July 25, 2022
- Address
- 1 Becton Dr, N/A, Franklin Lakes, NJ, 07417-1815, United States
Description
BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)
Reason
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
Code Info
All lots
Distribution
Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
Quantity
3,956,900