FDA Enforcement
Class I
Ongoing
NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)
Recall: Z-1138-2023
·
Reported March 1, 2023
Enforcement
- Recall Number
- Z-1138-2023
- Event ID
- 91538
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Draeger Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 1, 2023
- Initiation Date
- January 23, 2023
- Classification Date
- February 23, 2023
- Address
- 3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States
Description
NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)
Reason
The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
Code Info
All Lot numbers. UDI-DI: 04048675040552 (MP01579); 04048675040569 (MP01580); 04048675040538 (MP01581)
Distribution
US Nationwide Distribution
Quantity
893 units