FDA Enforcement Class II Terminated

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

Recall: Z-1138-2019 · Reported April 24, 2019

Enforcement

Recall Number
Z-1138-2019
Event ID
82322
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Topcon Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 24, 2019
Initiation Date
March 6, 2019
Classification Date
April 12, 2019
Termination Date
July 17, 2019
Address
111 Bauer Dr, N/A, Oakland, NJ, 07436-3123, United States

Description

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

Reason

When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.

Code Info

System Version 1.53

Distribution

Nationwide

Quantity

450 units