VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.
Enforcement
- Recall Number
- Z-1136-2021
- Event ID
- 87200
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ortho-Clinical Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 3, 2021
- Initiation Date
- January 14, 2021
- Classification Date
- February 19, 2021
- Termination Date
- December 27, 2022
- Address
- 100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States
Description
VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.
A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.
Lot 413420 Model(s): VITROS 250, VITROS 350, VITROS ECi, VITROS 5,1 FS, VITROS 3600, VITROS 4600, VITROS 5600, VITROS XT7600.
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada.
4,062 cartons of 1,000 tips (Totaling 4,062,000 tips)