SIEMENS Biograph mCT: a) Flow 64-4R Seismic, b) S(40)-3R Seismic, c) S(64)-3R Seismic, d) S(40)-4R Seismic, e) S(64)-4R Seismic, f) S(20)-3R Mobile
Enforcement
- Recall Number
- Z-1135-2023
- Event ID
- 91518
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 22, 2023
- Initiation Date
- January 16, 2023
- Classification Date
- February 13, 2023
- Address
- 2501 Barrington Rd, N/A, Hoffman Estates, IL, 60192-2061, United States
Description
SIEMENS Biograph mCT: a) Flow 64-4R Seismic, b) S(40)-3R Seismic, c) S(64)-3R Seismic, d) S(40)-4R Seismic, e) S(64)-4R Seismic, f) S(20)-3R Mobile
This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.
a) Flow 64-4R Seismic, UDI/DI 405686900142560047, Material Number 10529161, Serial Number 60047; b) S(40)-3R Seismic, UDI/DI 405686900131960054, Material Number 10248668, Serial Number 60054; c) S(64)-3R Seismic, UDI/DI 405686900133360091, Material Number 10248669, Serial Number 60091; d) S(40)-4R Seismic, UDI/DI 405686900132660047, Material Number 10248671, Serial Number 60047; e) S(64)-4R Seismic, UDI/DI 405686900134060076, Material Number 10248672, Serial Number 60076; f) S(20)-3R Mobile, UDI/DI 405686900129660028, , Material Number 10534159, Serial Number 60028
Worldwide distribution - US Nationwide and the countries of Japan, Chile, Australia.
6 units