FDA Enforcement
Class II
Ongoing
iTotal Identity Impactor Handle, Model No. 1080-212/ED-07827 REV AE
Recall: Z-1133-2023
·
Reported February 22, 2023
Enforcement
- Recall Number
- Z-1133-2023
- Event ID
- 91376
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Conformis, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 22, 2023
- Initiation Date
- November 14, 2022
- Classification Date
- February 13, 2023
- Address
- 600 Technology Park Dr Ste 3, N/A, Billerica, MA, 01821-4154, United States
Description
iTotal Identity Impactor Handle, Model No. 1080-212/ED-07827 REV AE
Reason
Loctite thread locker (epoxy) may not have been applied to the screw that is part of the handle. In the absence of epoxy, the screw may loosen and fall into the patient during a procedure.
Code Info
UDI-DI 00810933030582 Lot D221301
Distribution
US Nationwide distribution in the states of AL, AZ, CO, FL, GA, KS, KY, LA, MA, MD, MS, NE, NH, NV, NY, OH, TN, WI, WY.
Quantity
70