FDA Enforcement Class II Terminated

Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

Recall: Z-1130-2015 · Reported February 25, 2015

Enforcement

Recall Number
Z-1130-2015
Event ID
70352
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Haemonetics Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 25, 2015
Initiation Date
January 26, 2015
Classification Date
February 19, 2015
Termination Date
June 5, 2015
Address
400 Wood Rd, Braintree, MA, 02184-2412, United States

Description

Haemonetics TEG Hemostasis System Level II Control, Catalog number: 8002. Level II Control is a biological control for use with the Thrombelastograph (TEG) Hemostasis System.

Reason

Specific lots of Level II Control Kits produce high Maximum Amplitude (MA) results, which are outside of the manufacturer's expected Quality Control range for MA parameter.

Code Info

Lot numbers: HMO 9200 HMO 9217 HMO 9226 HMO 9232 HMO 9276 HMO 9285 HMO 9292 Exp. Date 1/2017

Distribution

Worldwide Distribution -- US, Canada, Czech Republic, Denmark, Great Britain, and Sweden.

Quantity

4,143 kits