LifeScan brand OneTouch¿ Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels.
Enforcement
- Recall Number
- Z-1130-2013
- Event ID
- 64478
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Lifescan Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 24, 2013
- Initiation Date
- February 25, 2013
- Classification Date
- April 17, 2013
- Termination Date
- October 23, 2013
- Address
- 1000 Gibraltar Dr, N/A, Milpitas, CA, 95035-6301, United States
Description
LifeScan brand OneTouch¿ Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels.
When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.
2AA2101, Exp. Date: 9-30-2014; 2AA2102, Exp. Date: 10-31-2-14
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.
132,301 units total