FDA Enforcement Class II Terminated

ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.

Recall: Z-1126-2019 · Reported April 17, 2019

Enforcement

Recall Number
Z-1126-2019
Event ID
82497
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ulrich medical USA Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 17, 2019
Initiation Date
March 8, 2019
Classification Date
April 11, 2019
Termination Date
July 10, 2020
Address
18221 Edison Ave, N/A, Chesterfield, MO, 63005-3703, United States

Description

ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.

Reason

Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.

Code Info

Model #s CS 7130 Serial #s 538436-02_70 (69 total) UDI Description 4052536123704, Model #s CS 7134 Serial #s 538438-02_70 (69 total) UDI Description 4052536123728, Model #s CS 7146 Serial #s 04-14-022-18_23 (6 total) UDI Description4052536078301, Model #s CS 7148 Serial #s 11-12-035-01_09, 03-12-006-24, and 03-12-006-25 (11 total) UDI Description 4052536078318, and Model #s CS 8049 Serial #s J04192-31_40, J04192-056, J04192-064, H12015-031_6, and J01276-031_3, (21 total) UDI Description 4052536022076.

Distribution

AZ, IL, LA, MI, MO, NV, OH, PA, TN, and TX.

Quantity

25