FDA Enforcement Class II Terminated

Covidien Nellcor SpO2 Forehead Sensor Item Code: RS10; The Covidien Nellcor¿ Forehead SpO2Sensor, model RS10, is indicated for use when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required for patients weighing more than 40 kg.

Recall: Z-1125-2017 · Reported February 8, 2017

Enforcement

Recall Number
Z-1125-2017
Event ID
76062
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Covidien LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 8, 2017
Initiation Date
December 16, 2016
Classification Date
February 2, 2017
Termination Date
December 4, 2018
Address
60 Middletown Ave, N/A, North Haven, CT, 06473-3908, United States

Description

Covidien Nellcor SpO2 Forehead Sensor Item Code: RS10; The Covidien Nellcor¿ Forehead SpO2Sensor, model RS10, is indicated for use when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required for patients weighing more than 40 kg.

Reason

Labeling error: The label indicates the product is latex free, but the headband component included with the forehead sensor does contain latex.

Code Info

Lot Numbers: 161510095H, 161620089H, 161690203H, 161760132H, 161900091H,161970090H, 162040107H, 162110103H, 162320191H, 162390212H

Distribution

Nationwide. Foreign: Belgium Germany France Netherlands

Quantity

7453 Units