FDA Enforcement Class II Terminated

Getinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector intended for washing, disinfecting and drying of surgical instruments, containers, operating theatre shoes and operating theatre accessories normally used in hospitals and central disinfection facilities. The disinfected items must be further processed through terminal sterilization before use in any surgical procedure.

Recall: Z-1124-2021 · Reported February 24, 2021

Enforcement

Recall Number
Z-1124-2021
Event ID
87155
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Getinge Group Logistics America, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 24, 2021
Initiation Date
December 21, 2020
Classification Date
February 16, 2021
Termination Date
July 20, 2022
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

Getinge CM320 Series Washer Disinfector - Product Usage: A multi chamber washer disinfector intended for washing, disinfecting and drying of surgical instruments, containers, operating theatre shoes and operating theatre accessories normally used in hospitals and central disinfection facilities. The disinfected items must be further processed through terminal sterilization before use in any surgical procedure.

Reason

The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.

Code Info

Product Code/Part Number: CM320 4 WUWD, CM320 3 WWD, CM320K661, CM320K671, CM320K689, CM320K706, CM320K730, CM320WU, CM320WU K981, CM320WUWD. Serial Number: 31253, 31260, 31299, 31286, W50026718, W50027250, W50029991, W50034570, W50039234, WAA070170, WAA081312, WAA062340, WAA062341, WAA070619, WAA085063, WAA089556, WAA093936, WAA094247, WAA070618. UDI Code: 07340153700178 GETINGE CM320WO CM320 2 CHAMBER WO 07340153700185 GETINGE CM320WU CM320 2 CHAMBER WU 07340153700192 GET/NGE CM320WUWO CM320 4 CHAMBER WUWO 07340153700208 GETINGE CM320WWO CM320 3 CHAMBER WWO 07340153700215 GETINGE CM320WWWO CM320 4 CHAMBER WWWO 07340153710153 GETINGE CM320WOK966 CM320 2 CHAMBER WO

Distribution

Worldwide distribution - US Nationwide distribution including in the states of ND, PA, NC, VA, NE, MD, NY, MN and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Norway, Slovakia, Slovenia, South Korea, Sweden, Japan, China, Netherland, Italy, Kuwait and Czech Republic.

Quantity

189 units total (Domestic 19 units, Foreign 170 units)