FDA Enforcement
Class II
Terminated
Exprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-065, 495-00-075, 495-00-085, 495-01-065, 495-01-075, 495-01-085
Recall: Z-1123-2019
·
Reported April 17, 2019
Enforcement
- Recall Number
- Z-1123-2019
- Event ID
- 82379
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Encore Medical, Lp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 17, 2019
- Initiation Date
- September 21, 2018
- Classification Date
- April 10, 2019
- Termination Date
- September 16, 2020
- Address
- 9800 Metric Blvd, N/A, Austin, TX, 78758-5445, United States
Description
Exprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-065, 495-00-075, 495-00-085, 495-01-065, 495-01-075, 495-01-085
Reason
Complaints regarding loose proximal body bolts or bolts that were found through x-ray as sitting above the proximal body and not properly seated. Patient risks may include: pain, dissociation, and possible revision surgery
Code Info
All lots.
Distribution
US: TN, MN, TX, CO, IL, IN, OH, SC, MS, MI, RI, PA, ME, NY, WA, MO, AR, OK, LA, AL, FL, GA, UT, AZ, CA, NV,.
Quantity
22