FDA Enforcement Class II Ongoing

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Recall: Z-1120-2026 · Reported January 28, 2026

Enforcement

Recall Number
Z-1120-2026
Event ID
98128
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 28, 2026
Initiation Date
December 15, 2025
Classification Date
January 16, 2026
Address
600 N Field Dr, Lake Forest, IL, 60045-4835, United States

Description

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Reason

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Code Info

400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0

Distribution

Worldwide distribution - US Nationwide and the country of Philippines.