FDA Enforcement Class II Terminated

Product 22 consists of all product under product code: JDI and same usage: Item no: 811400218 CPT 12/14 COCR REVISION S 811400318 CPT 12/14 COCR REVISION S 811400420 CPT 12/14 COCR REVISION S 811400423 CPT 12/14 COCR REVISION S 811400426 CPT 12/14 COCR REVISION S 811400010 CPT 12/14 COCR SIZE 0 EXT 811400000 CPT 12/14 COCR SIZE 0 STD 811400110 CPT 12/14 COCR SIZE 1 EXT 811400100 CPT 12/14 COCR SIZE 1 STD 811400210 CPT 12/14 COCR SIZE 2 EXT 811400200 CPT 12/14 COCR SIZE 2 STD 811400230 CPT 12/14 COCR SIZE 2 XEX 811400310 CPT 12/14 COCR SIZE 3 EXT 811400300 CPT 12/14 COCR SIZE 3 STD 811400330 CPT 12/14 COCR SIZE 3 XEX 811400410 CPT 12/14 COCR SIZE 4 EXT 811400400 CPT 12/14 COCR SIZE 4 STD 811400430 CPT 12/14 COCR SIZE 4 XEX 811400510 CPT 12/14 COCR SIZE 5 EXT 811400500 CPT 12/14 COCR SIZE 5 STD 811400530 CPT 12/14 COCR SIZE 5 XEX For use in total or hemi hip arthroplasty

Recall: Z-1120-2016 · Reported March 23, 2016

Enforcement

Recall Number
Z-1120-2016
Event ID
72900
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Manufacturing B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 23, 2016
Initiation Date
January 11, 2016
Classification Date
March 12, 2016
Termination Date
April 13, 2017
Address
Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, 00715, United States

Description

Product 22 consists of all product under product code: JDI and same usage: Item no: 811400218 CPT 12/14 COCR REVISION S 811400318 CPT 12/14 COCR REVISION S 811400420 CPT 12/14 COCR REVISION S 811400423 CPT 12/14 COCR REVISION S 811400426 CPT 12/14 COCR REVISION S 811400010 CPT 12/14 COCR SIZE 0 EXT 811400000 CPT 12/14 COCR SIZE 0 STD 811400110 CPT 12/14 COCR SIZE 1 EXT 811400100 CPT 12/14 COCR SIZE 1 STD 811400210 CPT 12/14 COCR SIZE 2 EXT 811400200 CPT 12/14 COCR SIZE 2 STD 811400230 CPT 12/14 COCR SIZE 2 XEX 811400310 CPT 12/14 COCR SIZE 3 EXT 811400300 CPT 12/14 COCR SIZE 3 STD 811400330 CPT 12/14 COCR SIZE 3 XEX 811400410 CPT 12/14 COCR SIZE 4 EXT 811400400 CPT 12/14 COCR SIZE 4 STD 811400430 CPT 12/14 COCR SIZE 4 XEX 811400510 CPT 12/14 COCR SIZE 5 EXT 811400500 CPT 12/14 COCR SIZE 5 STD 811400530 CPT 12/14 COCR SIZE 5 XEX For use in total or hemi hip arthroplasty

Reason

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Code Info

lot no.: 61907682 62038367 62106797 62161479 62240743 62268577 62344683 62030932 62106796 62148557 62206538 62240742 62303507 62352823

Distribution

US Nationwide Distribution

Quantity