ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)
Enforcement
- Recall Number
- Z-1119-2026
- Event ID
- 98128
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- ICU Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 28, 2026
- Initiation Date
- December 15, 2025
- Classification Date
- January 16, 2026
- Address
- 600 N Field Dr, Lake Forest, IL, 60045-4835, United States
Description
ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
400020403 (US), Software Versions: 1.2.2; 400021003 (Philippines), Software Versions: 1.2.0
Worldwide distribution - US Nationwide and the country of Philippines.