Clinical Chemistry Phenobarbital, list number 1E08
Enforcement
- Recall Number
- Z-1118-2013
- Event ID
- 63945
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Laboratories, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 24, 2013
- Initiation Date
- November 16, 2012
- Classification Date
- April 15, 2013
- Termination Date
- June 18, 2013
- Address
- 1921 Hurd Dr, N/A, Irving, TX, 75038-4313, United States
Description
Clinical Chemistry Phenobarbital, list number 1E08
Abbott Diagnostics is recalling Clinical Chemistry Phenobarbital reagent lots 62299UN12 and 85773UN12 due to imprecision issues which increase with the age of the reagent lot. The imprecision of the Phenobarbital assay is d 7% Total CV%; however, internal testing confirmed CVs up to approximately 12% for in-date reagent lots over five months old. Although lot 52803UN12 is now expired, it exhibit
Lot Numbers 52803UN12, 62299UN12, and 85773UN12
Worldwide Distribution - USA (nationwide) and Internationally to Internationally distributed to Argentina, Australia, Austria, Bahamas, Brunei, Canada, Chile, Curacao, Denmark, France, Germany, Greece, Ireland, Italy, Lebanon, Mexico, Netherlands, Norway, Oman, Paraguay, Portugal, Puerto Rico, Russia, Saudi Arabia, Serbia, South Korea, Spain, Switzerland, Turkey, and United Kingdom.
1289