Medtronic PS Medical Cerebro Spinal Fluid (CSF)-Flow Control Valves for the following CFNs: 22011 L- SHUNT 22011 L INTL ONLY ULTRA SM LOW 22011 LL - SHUNT 22011 LL INTL ONLY ULTRA SM L-L 22011 M - SHUNT 22011 M INTL ONLY ULTRA SM MED 22017 B-L - SHUNT 22017 B-L ASSY US PER 90CM SM LOW 22017 B-M - SHUNT 22017 B-M ASSY US PER 90CM SM MED 22017B-LL - SHUNT 22017B-LL ASSY US PER 90CM SM LL 23093 - VALVE 23093 CONT. REG. LOW-LOW 27102 - VALVE 27102 CONTOURED SML LOW LOW 27553 M - VALVE 27553 M CSF-FLOW CNTRL SM EXTR MP 27643M - SHUNT 27643M ASSY SM PCATH130CM MED 27683M - SHUNT 27683M RG SNP PCATH90 PLI STD OE 42312 - VALVE 42312 SM CONTOURED LOW PRES IMP 42314- VALVE 42314 SM CONTOURED MED PRESS IMP 42316 - VALVE 42316 SM CONTOURED HIGH PRESS 42322 - VALVE 42322 CONTOURED REG LOW PRESS IMP 42324 - VALVE 42324 CONTOURED REG MED PRESS IMP 42326 - VALVE 42326 CONT HIGH PRESS IMP 42410 - VALVE 42410 ULTRA SMALL LOW-LOW 42414 - VALVE 42414 ULTRA SMALL MEDIUM PRESS IMP 42532 - VALVE 42532 BURR HOLE 12MM LOW PRESS IMP 42534 - VALVE 42534 BURR HOLE 12MM MED PRESS IMP 42536 - VALVE 42536 BURR HOLE 12MM HI PRESS IMP 42542 - VALVE 42542 BURR HOLE 16 MM LOW IMP 42544 - VALVE 42544 BURR HOLE 16MM MED PRESS IMP 42546 - VALVE 42546 BURR HOLE 16MM HI PRESS IMP 46564 - SHUNT 46564 CSF SNAP ASSY US MEDIUM 46622 - SHUNT 46622 CSF ASSY SMALL LOW CONT C/P 46624 - SHUNT 46624 CSF ASSY SMALL MED CONT C/P 46626 - SHUNT 46626 CSF ASSY SMALL HIGH CONT C/P 46642 - SHUNT 46642 ASSY REG LOW C/P-90CM 46644 - SHUNT 46644 ASSY REG MED C/P-90CM 46646 - SHUNT 46646 ASSY REG HIGH C/P-90CM 9003 A - KIT 9003 A INTL ONLY SHUNT CONT SM LO 9003 B - KIT 9003 B INTL ONLY SHUNT CONT SM ME 9003 C - KIT 9003 C INTL ONLY SHUNT CONT SM HI 9003 D - KIT 9003 D SHUNT INTL ONLY CONT REG L 9003 E - KIT 9003 E SHUNT INTL ONLY CONT REG M 9003 F - KIT 9003 F SHUNT INTL ONLY CONT REG H 9040 A - KIT 9040 A SHUNT INTL ONLY BH 12MM LOW 9040 B - KIT 9040 B SHUNT INTL ONLY BH 12MM MED 9040 C - KIT 9040 C SHUNT INTL ONLY BH 12MM HIGH 9040 D - KIT 9040 DSHUNT INTL ONLY BH 16MM LOW 9040 E - KIT 9040 E SHUNT INTL ONLY BH 16MM MED 9040 F - KIT 9040 A SHUNT INTL ONLY BH 12MM LOW 92322 - VALVE 92322 CSF CONT REG LOW BIO 92324 - VALVE 92324 CSF CONT REG MED BIO
Enforcement
- Recall Number
- Z-1117-2021
- Event ID
- 87157
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medtronic Neurosurgery
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 24, 2021
- Initiation Date
- January 11, 2021
- Classification Date
- February 12, 2021
- Address
- 5290 California Ave, N/A, Irvine, CA, 92617-3073, United States
Description
Medtronic PS Medical Cerebro Spinal Fluid (CSF)-Flow Control Valves for the following CFNs: 22011 L- SHUNT 22011 L INTL ONLY ULTRA SM LOW 22011 LL - SHUNT 22011 LL INTL ONLY ULTRA SM L-L 22011 M - SHUNT 22011 M INTL ONLY ULTRA SM MED 22017 B-L - SHUNT 22017 B-L ASSY US PER 90CM SM LOW 22017 B-M - SHUNT 22017 B-M ASSY US PER 90CM SM MED 22017B-LL - SHUNT 22017B-LL ASSY US PER 90CM SM LL 23093 - VALVE 23093 CONT. REG. LOW-LOW 27102 - VALVE 27102 CONTOURED SML LOW LOW 27553 M - VALVE 27553 M CSF-FLOW CNTRL SM EXTR MP 27643M - SHUNT 27643M ASSY SM PCATH130CM MED 27683M - SHUNT 27683M RG SNP PCATH90 PLI STD OE 42312 - VALVE 42312 SM CONTOURED LOW PRES IMP 42314- VALVE 42314 SM CONTOURED MED PRESS IMP 42316 - VALVE 42316 SM CONTOURED HIGH PRESS 42322 - VALVE 42322 CONTOURED REG LOW PRESS IMP 42324 - VALVE 42324 CONTOURED REG MED PRESS IMP 42326 - VALVE 42326 CONT HIGH PRESS IMP 42410 - VALVE 42410 ULTRA SMALL LOW-LOW 42414 - VALVE 42414 ULTRA SMALL MEDIUM PRESS IMP 42532 - VALVE 42532 BURR HOLE 12MM LOW PRESS IMP 42534 - VALVE 42534 BURR HOLE 12MM MED PRESS IMP 42536 - VALVE 42536 BURR HOLE 12MM HI PRESS IMP 42542 - VALVE 42542 BURR HOLE 16 MM LOW IMP 42544 - VALVE 42544 BURR HOLE 16MM MED PRESS IMP 42546 - VALVE 42546 BURR HOLE 16MM HI PRESS IMP 46564 - SHUNT 46564 CSF SNAP ASSY US MEDIUM 46622 - SHUNT 46622 CSF ASSY SMALL LOW CONT C/P 46624 - SHUNT 46624 CSF ASSY SMALL MED CONT C/P 46626 - SHUNT 46626 CSF ASSY SMALL HIGH CONT C/P 46642 - SHUNT 46642 ASSY REG LOW C/P-90CM 46644 - SHUNT 46644 ASSY REG MED C/P-90CM 46646 - SHUNT 46646 ASSY REG HIGH C/P-90CM 9003 A - KIT 9003 A INTL ONLY SHUNT CONT SM LO 9003 B - KIT 9003 B INTL ONLY SHUNT CONT SM ME 9003 C - KIT 9003 C INTL ONLY SHUNT CONT SM HI 9003 D - KIT 9003 D SHUNT INTL ONLY CONT REG L 9003 E - KIT 9003 E SHUNT INTL ONLY CONT REG M 9003 F - KIT 9003 F SHUNT INTL ONLY CONT REG H 9040 A - KIT 9040 A SHUNT INTL ONLY BH 12MM LOW 9040 B - KIT 9040 B SHUNT INTL ONLY BH 12MM MED 9040 C - KIT 9040 C SHUNT INTL ONLY BH 12MM HIGH 9040 D - KIT 9040 DSHUNT INTL ONLY BH 16MM LOW 9040 E - KIT 9040 E SHUNT INTL ONLY BH 16MM MED 9040 F - KIT 9040 A SHUNT INTL ONLY BH 12MM LOW 92322 - VALVE 92322 CSF CONT REG LOW BIO 92324 - VALVE 92324 CSF CONT REG MED BIO
Potential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus
Lot numbers for the affected CFNs: 0219278853, 0219278854, 0220099072, 0219278855, 0219278856, 0219768090, 0219768091, 0219798970, 0219798971, 0219798972, 0220010048, 0220112693, 0220141392, 0219802495, 0219802494, 0219583129, 0220286100, 0218796592, 0219952709, 0218806424, 0219316930, 0220280248, 0220286709, 0219480202, 0219802492, 0218045303, 0218045304, 0218045527, 0218408997, 0218408998, 0218408999, 0218590082, 0218721471, 0218721472, 0219234579, 0219497114, 0219648109, 0219978722, 0220100564, 0220444728, 0217286212, 0217286213, 0217286214, 0217706616, 0217706621, 0217706622, 0217706623, 0217962420, 0217962421, 0217971628, 0217971633, 0218409003, 0218409004, 0218409005, 0218590077, 0218590078, 0218590079, 0218590080, 0218590081, 0218720196, 0218720202, 0218972089, 0218972090, 0219234580, 0219312365, 0219375930, 0219761087, 0220122285, 0220124886, 0220124887, 0220124888, 0220272309, 0220444084, 0220444085, 0220444726, 0217286211, 0217706612, 0217962419, 0217971634, 0217971635, 0218424703, 0218424704, 0218590071, 0217286216, 0217962409, 0217962418, 0217971637, 0217971638, 0218424701, 0218424702, 0217962422, 0217962423, 0217962424, 0217962425, 0218402220, 0218408987, 0218408988, 0218408989, 0218408990, 0218408991, 0218408992, 0218408993, 0218408994, 0218408996, 0218590073, 0218590074, 0218590075, 0218590076, 0218720244, 0218720287, 0218720289, 0219234581, 0219234582, 0219320159, 0219320160, 0219320162, 0219320163, 0219574223, 0219761097, 0219910255, 0219910256, 0220324658, 0220530229, 0218590084, 0218590085, 0219312364, 0219978742, 0220444083, 0219277363, 0219278519, 0219278520, 0219278516, 0219278517, 0219374313, 0219374314, 0219733266, 0219733267, 0218605545, 0218605686, 0218605539, 0218605540, 0218605541, 0218605542, 0218605543, 0218605544, 0219375125, 0219375926, 0219375927, 0219375928, 0219375929, 0219438310, 0219761086, 0219910252, 0219910253, 0220444735, 0218605538, 0218605536, 0218605537, 0220280189, 0218605688, 0218605689, 0218605690, 0218605691, 0218978999, 0218979001, 0218979002, 0218979003, 0218979005, 0218979086, 0218979087, 0219123064, 0219123586, 0219123588, 0219123589, 0219422860, 0219422861, 0219465702, 0219465703, 0219761093, 0219761094, 0219761095, 0219761096, 0220272311, 0218605687, 0219579789, 0219894795, 0219279151, 0220288369, 0219278874, 0219894797, 0220201981, 0220602456, 0220285339, 0219279160, 0219278877, 0219278878, 0219278879, 0219278880, 0219694364, 0220075413, 0220075414, 0219354851, 0219354852, 0219354853, 0219354854, 0220114768, 0220398628, 0219278737, 0219278738, 0219278739, 0219278740, 0219708098, 0219708099, 0219851768, 0219851769, 0219851770, 0219981926, 0220053812, 0220053813, 0220081704, 0220085756, 0220085757, 0220099077, 0220099078, 0220134024, 0220134025, 0220156462, 0220156463, 0220173989, 0220285673, 0220352833, 0220398644, 0219278736, 0220393229, 0219278734, 0219278735, 0220011262, 0219278731, 0219278733, 0219769315, 0219769316, 0219851771, 0219851772, 0219851773, 0220305451, 0220622330, 0220680023, 0219354855, 0219927060, 0220546836, 0219278730, 0220011259, 0220397733, 0219127889, 0219128109, 0219128111, 0219128112, 0219128113, 0219128115, 0219128117, 0219128118, 0219128119, 0219128120, 0219128121, 0219917405, 0220009922, 0220009923, 0220040136, 0220059298, 0220059299, 0220099075, 0220099076, 0220102737, 0220118172, 0220156744, 0220156745, 0220201972, 0220201979, 0220217393, 0220249127, 0220397688, 0220397689, 0220397691, 0220397699, 0220397703, 0220400359, 0220702309, 0219969972, 0219272921, 0219950183, 0219127462, 0219127463, 0219127464, 0219127465, 0219127886, 0219127887, 0219127890, 0219127891, 0219127892, 0219127893, 0219127894, 0219127895, 0219127896, 0219127897, 0219127898, 0219698790, 0219698792, 0219721377, 0219721378, 0219721379, 0219721380, 0219721381, 0219917404, 0219988521, 0220075419, 0220141390, 0220141391, 0220141405, 0220141706, 0220156743, 0220173988, 0220249134, 0220255058, 0220262090, 0220285669, 0220285670, 0220285672, 0220301203, 0220324602, 0220330904, 0220360618, 0220637047, 0219278729, 0219698786, 0220522999, 0219067893, 0219067894, 0220400350,
Worldwide distribution. US nationwide, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Plurinational State Of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Islamic Republic Of Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Republic Of Korea, Kuwait, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macao, Malaysia, Malta, Mauritania, Mexico, Montenegro, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, and Yemen
20,817 units