FDA Enforcement Class I Terminated

Edwards Lifesciences, Fogarty Dilation Atrioseptostomy Catheter, Sterile EO, Rx Only, REF: 830705F Product Usage: The Fogarty dilation atrioseptostomy catheter is designed for enlarging interatrial openings. In most pediatric cardiology centers, balloon atrioseptostomy is an accepted technique for palliation of several congenital cardiac defects. Description of Use Balloon atrioseptostomy can be performed in conjunction with diagnostic cardiac catheterization or under echocardiographic guidance and has been carried out after the diagnosis of several congenital cardiac defects including but not limited to: transposition of the great arteries, total anomalous pulmonary venous drainage without pulmonary obstruction, tricuspid atresia, and pulmonary atresia with intact ventricular septum.

Recall: Z-1116-2019 · Reported May 1, 2019

Enforcement

Recall Number
Z-1116-2019
Event ID
82455
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Edwards Lifesciences, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 1, 2019
Initiation Date
March 28, 2019
Classification Date
April 25, 2019
Termination Date
November 1, 2023
Address
1 Edwards Way, N/A, Irvine, CA, 92614-5688, United States

Description

Edwards Lifesciences, Fogarty Dilation Atrioseptostomy Catheter, Sterile EO, Rx Only, REF: 830705F Product Usage: The Fogarty dilation atrioseptostomy catheter is designed for enlarging interatrial openings. In most pediatric cardiology centers, balloon atrioseptostomy is an accepted technique for palliation of several congenital cardiac defects. Description of Use Balloon atrioseptostomy can be performed in conjunction with diagnostic cardiac catheterization or under echocardiographic guidance and has been carried out after the diagnosis of several congenital cardiac defects including but not limited to: transposition of the great arteries, total anomalous pulmonary venous drainage without pulmonary obstruction, tricuspid atresia, and pulmonary atresia with intact ventricular septum.

Reason

Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of balloon fragmentation or detachment.

Code Info

All lots within 15 months expiry from 03/01/2019

Distribution

Worldwide Distribution - US Nationwide in the states of : IA, FL, MD, CA, NY, TX, CT, NC, OR, LA, UT,AK, WA, AZ, SD,CO,IN, KY, WV, TN, MI, AL, MA, VA, PA, MO, IL, GA, PR, MN, AR, OK, WI. Foreign (OUS): CL, AR, CZ, SG, BR, ID, ES, IE, GB, AT, CA, AM, HK, AU, ZA, FI, KR, MY, TH, NL, FR, IT, GR, DK, SE, DE, TR, QA, NZ, SA, PS, PT, JP, CH, IL, AE, BE, EE

Quantity

1093