FDA Enforcement Class II Terminated

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Recall: Z-1116-2015 · Reported February 25, 2015

Enforcement

Recall Number
Z-1116-2015
Event ID
70492
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 25, 2015
Initiation Date
February 6, 2015
Classification Date
February 13, 2015
Termination Date
October 17, 2016
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586, United States

Description

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Reason

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Code Info

HX2 Temperature Management System, 120V Catalog # 809810 serial numbers: 00016, 0001  1116 HX2 Temperature Management System, 100V Catalog # 812231 serial numbers: 0001, 1002, PT0003, PT0004 HX2 Temperature Management System, 240V Catalog # 812240 serial numbers: 0001, 0002 HX2 Temperature Management System, Trial Kit Catalog # 809810TK serial numbers: 1001  1042

Distribution

US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.

Quantity

103 units