FDA Enforcement Class II Ongoing

VITROS Immunodiagnostic Products Prolactin Reagent Pack Product Code: 1849793

Recall: Z-1114-2022 · Reported May 25, 2022

Enforcement

Recall Number
Z-1114-2022
Event ID
90047
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 25, 2022
Initiation Date
March 25, 2022
Classification Date
May 18, 2022
Address
1001 Us Highway 202, N/A, Raritan, NJ, 08869-1424, United States

Description

VITROS Immunodiagnostic Products Prolactin Reagent Pack Product Code: 1849793

Reason

Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.

Code Info

Until further notice, future lots will also be affected. UDI: 10758750000111 Affected Lot Number/ Expiration Date: 2840 13-MAY-2022; 2850 08-JUN-2022; 2860 02-JUL-2022; 2870 17-JUL-2022; 2875 17-JUL-2022; 2880 14-SEP-2022; 2891 28-SEP-2022; 2900 29-OCT-2022;

Distribution

Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom

Quantity

1767 units total: 1521 US; 246 OUS