FDA Enforcement Class II Terminated

CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").

Recall: Z-1114-2013 · Reported April 24, 2013

Enforcement

Recall Number
Z-1114-2013
Event ID
64460
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Diagnostics Operations, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 24, 2013
Initiation Date
March 6, 2013
Classification Date
April 15, 2013
Termination Date
January 10, 2018
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").

Reason

Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.

Code Info

catalog number 05530199160

Distribution

Nationwide Distribution.

Quantity

4467 both devices