FDA Enforcement Class II Ongoing

Philips Azurion system; Software Version Number: R3.1;

Recall: Z-1113-2026 · Reported January 28, 2026

Enforcement

Recall Number
Z-1113-2026
Event ID
98108
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 28, 2026
Initiation Date
December 12, 2025
Classification Date
January 16, 2026
Address
Veenpluis 4-6, Best, N/A, Netherlands

Description

Philips Azurion system; Software Version Number: R3.1;

Reason

Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

Code Info

Software Version Number: R3.1; (1) Model Number: 722221; System Product Name: Azurion 3 M12; UDI-DI: 00884838099203; Serial Numbers: (2) Model Number: 722222; System Product Name: Azurion 3 M15; UDI-DI: 00884838099210; Serial Numbers: (3) Model Number: 722223; System Product Name: Azurion 7 M12; UDI-DI: 00884838099241; Serial Numbers: (4) Model Number: 722224; System Product Name: Azurion 7 M20; UDI-DI: 00884838099258; Serial Numbers: (5) Model Number: 722225; System Product Name: Azurion 7 B12; UDI-DI: 00884838099265; Serial Numbers: (6) Model Number: 722226; System Product Name: Azurion 7 B20; UDI-DI: 00884838099272; Serial Numbers: (7) Model Number: 722227; System Product Name: Azurion 5 M12; UDI-DI: 00884838099227; Serial Numbers: (8) Model Number: 722228; System Product Name: Azurion 5 M20; UDI-DI: 00884838099234; Serial Numbers: (9) Model Number: 722229; System Product Name: Azurion 3 M12; UDI-DI: 00884838116726; Serial Numbers: 5, 9, 32, 40, 35, 31, 33; (10) Model Number: 722230; System Product Name: Azurion 3 M15; UDI-DI: 00884838116733; Serial Numbers: 25, 54, 50, 52, 55, 30, 63, 60, 67; (11) Model Number: 722231; System Product Name: Azurion 5 M12; UDI-DI: 00884838116740; Serial Numbers: 62, 7, 8, 14, 17, 63, 6, 66, 59, 69, 56, 67, 48, 68, 80, 93, 87, 94, 81, 92, 18; (12) Model Number: 722232; System Product Name: Azurion 5 M20; UDI-DI: 00884838116757; Serial Numbers: 97, 39, 31, 100, 104, 93, 115, 113, 118, 119, 89, 54, 67, 7, 128, 86, 110, 132, 130, 131, 123, 137, 139, 117, 260; (13) Model Number: 722233; System Product Name: Azurion 7 M12; UDI-DI: 00884838116764; Serial Numbers: 239, 100, 233, 240, 176, 227, 223, 130, 161, 194, 243, 56, 193, 280, 75, 253, 156, 288, 72, 281, 304, 300, 105, 295, 312, 310, 285, 292, 290, 298, 297; (14) Model Number: 722234; System Product Name: Azurion 7 M20; UDI-DI: 00884838116771; Serial Numbers: 731, 727, 261, 473, 683, 722, 562, 750, 745, 766, 265, 701, 400, 310, 421, 314, 182, 579, 96, 763, 272, 612, 15, 844, 105, 702, 751, 384, 851, 869, 743, 337, 651, 455, 706, 312, 563, 709, 764, 871, 846, 278, 713, 802, 868, 277, 742, 184, 819, 759, 910, 928, 415, 414, 945, 961, 355, 933, 692, 920, 569, 939, 909, 553, 837, 856, 571, 756, 941, 570, 761, 981, 353, 888, 964, 820, 746, 736, 876, 599, 1017, 908, 487, 983, 495; (15) Model Number: 722235; System Product Name: Azurion 7 B12; UDI-DI: 00884838116788; Serial Numbers: 96, 79, 111, 100, 101, 38, 27, 92, 78, 118, 48, 131, 135, 22, 125, 103, 138, 139, 137; (16) Model Number: 722236; System Product Name: Azurion 7 B20; UDI-DI: 00884838116801; Serial Numbers: 241, 297, 198, 215, 305, 274, 299, 78, 279, 287, 318, 353, 324, 331, 293, 245, 314, 310, 317, 312, 77, 358, 206, 383, 380, 370, 377, 385, 405, 379, 328;

Distribution

International distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA;

Quantity

228 units (all OUS)