FDA Enforcement Class I Terminated

HUDSON RCI COMFORT FLO Corrugated Humidification System with Remote Temperature Port, REF 2416, humidifier nebulizer kit

Recall: Z-1113-2020 · Reported February 19, 2020

Enforcement

Recall Number
Z-1113-2020
Event ID
84588
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
TELEFLEX MEDICAL INC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 19, 2020
Initiation Date
December 13, 2019
Classification Date
February 7, 2020
Termination Date
April 30, 2024
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States

Description

HUDSON RCI COMFORT FLO Corrugated Humidification System with Remote Temperature Port, REF 2416, humidifier nebulizer kit

Reason

There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.

Code Info

GTIN 14026704659378 Lot Numbers: 74J1500186 74L1502160 74M1500351 74A1600591 74A1601240 74A1602237 74C1601017 74C1602179 74D1600258 74D1601234 74D1601806 74E1600055 74G1600015 74H1600056 74J1601477 74J1602933 74K1602258 74L1600988 74M1600529 74B1701595 74C1700695 74E1702180 74F1700032 74G1700703 74G1700591 74G1700804 74H1701221 74J1700696 74J1700697 74J1700698 74K1700304 74K1702510 74A1802332 74A1802413 74B1801277 74C1800029 74C1800491 74C1800455 74C1801364 74C1801891 74D1800076 74E1800783 74E1800782 74F1800029 74G1800017 74H1801292 74J1800059 74K1800024 74L1800063 74L1802366 74M1800652 74A1901353 74A1901830 74B1900649 74C1900439 74D1900088 74D1900181

Distribution

nationwide

Quantity

60467 devices