FDA Enforcement Class II Terminated

cobas¿ PCR Urine Kit 100 PKT IVD cobas¿ PCR Female Swab Kit 100 PKT IVD cobas¿ PCR media 100T IVD cobas¿ PCR Female Swab Kit 100 PKT JPN-IVD Roche Molecular Systems, Inc. Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 The cobas¿¿ PCR media serves as a nucleic acid stabilizing transport and storage medium for gynecological specimens and is contained in the cobas¿¿ PCR Female Swab Kit, cobas¿¿ PCR Urine kit, and cobas¿¿ PCR media 100 T kit. -The cobas¿¿ PCR Urine Sample Kit is used to collect and transport urine specimens. Use this collection kit only with the cobas¿¿ CT/NG Test. - The cobas¿¿ PCR Female Swab Sample Kit is used to collect and transport endocervical and vaginal swab specimens. Use this collection kit only with the cobas¿¿ CT/NG Test NOTE: This collection kit should not be used for collection of female urine specimens. o For endocervical swab specimens, collections are to be performed by a clinician. o For vaginal swab specimens, collections are to be performed by a clinician or by the patient (self-collection) -In US labeling, it indicates Self collection in a clinical setting. - In ex-US labeling, it indicates Clinician Instructed Self- Collection and Ensure the patient has read and understood the following self-collection instructions before providing a collection kit. - In some countries (e.g., UK), the sample collection kit is provided as self-collection for home / personal use, which may not be used in a controlled clinical setting and it is unknown if any instructions are provided. - The cobas¿¿ PCR media kit is used to stabilize and transport urine specimens. Use this kit only with the cobas¿¿ CT/NG Test.

Recall: Z-1111-2014 · Reported March 5, 2014

Enforcement

Recall Number
Z-1111-2014
Event ID
67441
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Molecular Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 5, 2014
Initiation Date
January 6, 2014
Classification Date
February 26, 2014
Termination Date
August 7, 2014
Address
1080 Us Highway 202 S, Branchburg, NJ, 08876-3733, United States

Description

cobas¿ PCR Urine Kit 100 PKT IVD cobas¿ PCR Female Swab Kit 100 PKT IVD cobas¿ PCR media 100T IVD cobas¿ PCR Female Swab Kit 100 PKT JPN-IVD Roche Molecular Systems, Inc. Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 The cobas¿¿ PCR media serves as a nucleic acid stabilizing transport and storage medium for gynecological specimens and is contained in the cobas¿¿ PCR Female Swab Kit, cobas¿¿ PCR Urine kit, and cobas¿¿ PCR media 100 T kit. -The cobas¿¿ PCR Urine Sample Kit is used to collect and transport urine specimens. Use this collection kit only with the cobas¿¿ CT/NG Test. - The cobas¿¿ PCR Female Swab Sample Kit is used to collect and transport endocervical and vaginal swab specimens. Use this collection kit only with the cobas¿¿ CT/NG Test NOTE: This collection kit should not be used for collection of female urine specimens. o For endocervical swab specimens, collections are to be performed by a clinician. o For vaginal swab specimens, collections are to be performed by a clinician or by the patient (self-collection) -In US labeling, it indicates Self collection in a clinical setting. - In ex-US labeling, it indicates Clinician Instructed Self- Collection and Ensure the patient has read and understood the following self-collection instructions before providing a collection kit. - In some countries (e.g., UK), the sample collection kit is provided as self-collection for home / personal use, which may not be used in a controlled clinical setting and it is unknown if any instructions are provided. - The cobas¿¿ PCR media kit is used to stabilize and transport urine specimens. Use this kit only with the cobas¿¿ CT/NG Test.

Reason

Numerous complaints have been filed that leaking cobas ¿ PCR media 4.3 mL IVD have been received from different lots of the cobas ¿ PCR Urine kits, cobas ¿ PCR Female Swab kits, and cobas ¿ PCR media kits.

Code Info

K110923 cobas PCR Female Swab Sample Kit 05170516190 Lot S01487 and earlier cobas PCR Urine Sample Kit 05170486190 Lot S01885 and earlier (except lot number S01884) IVD: 48 kit batches US: 31 kit batches JPN-IVD:6 kit batches

Distribution

Nationwide Distribution

Quantity

84 kit batches