FDA Enforcement Class II Terminated

Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.

Recall: Z-1109-2019 · Reported April 17, 2019

Enforcement

Recall Number
Z-1109-2019
Event ID
82384
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Elekta Limited
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 17, 2019
Initiation Date
March 12, 2019
Classification Date
April 8, 2019
Termination Date
November 3, 2023
Address
Linac House, Fleming Way, Crawley, N/A, N/A, United Kingdom

Description

Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.

Reason

Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.

Code Info

Unity Systems: 13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009,

Distribution

Worldwide Distribution - US Nationwide in the states of TX, NY, WI, and Denmark, Germany, Italy, Netherlands, Sweden, United Kingdom

Quantity

12 units