FDA Enforcement Class II Terminated

Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Recall: Z-1109-2017 · Reported February 8, 2017

Enforcement

Recall Number
Z-1109-2017
Event ID
76159
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Immuno-Mycologics, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 8, 2017
Initiation Date
December 27, 2016
Classification Date
January 27, 2017
Termination Date
March 8, 2017
Address
2700 Technology Pl, N/A, Norman, OK, 73071-1127, United States

Description

Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Reason

The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.

Code Info

161028133653JH, 161003111544KH, and 161007105036JG

Distribution

Nationwide throughout the US

Quantity

630 units