FDA Enforcement
Class II
Terminated
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Recall: Z-1109-2017
·
Reported February 8, 2017
Enforcement
- Recall Number
- Z-1109-2017
- Event ID
- 76159
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Immuno-Mycologics, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 8, 2017
- Initiation Date
- December 27, 2016
- Classification Date
- January 27, 2017
- Termination Date
- March 8, 2017
- Address
- 2700 Technology Pl, N/A, Norman, OK, 73071-1127, United States
Description
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Reason
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
Code Info
161028133653JH, 161003111544KH, and 161007105036JG
Distribution
Nationwide throughout the US
Quantity
630 units