FDA Enforcement Class II Ongoing

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No

Recall: Z-1108-2026 · Reported January 28, 2026

Enforcement

Recall Number
Z-1108-2026
Event ID
98117
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 28, 2026
Initiation Date
November 28, 2025
Classification Date
January 16, 2026
Address
Eugeniavagen 18c, Stockholm, N/A, Sweden

Description

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No

Reason

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Code Info

UDI: 0735000201076120240508, 0735000201080820241217 GTIN: 07350002010761, 07350002010808 Serial Numbers: 15.1.0.852, 15.2.0.107 Software Revision: RayStation 2024A SP1, RayStation 2024A SP2 Expiration Date: 2030-04-15

Distribution

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Quantity

0