Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
Enforcement
- Recall Number
- Z-1108-2026
- Event ID
- 98117
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 28, 2026
- Initiation Date
- November 28, 2025
- Classification Date
- January 16, 2026
- Address
- Eugeniavagen 18c, Stockholm, N/A, Sweden
Description
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
UDI: 0735000201076120240508, 0735000201080820241217 GTIN: 07350002010761, 07350002010808 Serial Numbers: 15.1.0.852, 15.2.0.107 Software Revision: RayStation 2024A SP1, RayStation 2024A SP2 Expiration Date: 2030-04-15
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
0