FDA Enforcement Class II Terminated

VITROS 5600 Integrated System, Product Code 6802413. Intended for use in the measurement of a variety of chemistry analytes.

Recall: Z-1108-2013 · Reported April 24, 2013

Enforcement

Recall Number
Z-1108-2013
Event ID
64577
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 24, 2013
Initiation Date
February 7, 2013
Classification Date
April 13, 2013
Termination Date
April 20, 2017
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

VITROS 5600 Integrated System, Product Code 6802413. Intended for use in the measurement of a variety of chemistry analytes.

Reason

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.

Code Info

Serial Numbers 56000118 through 56001583

Distribution

Worldwide Distribution- USA (nationwide) including Puerto Rico and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Republic of Panama, Singapore, Spain, and Venezuela.

Quantity

USA: 775 units, Foreign: 800 units