FDA Enforcement Class I Ongoing

MEDLINE Kits, trays, and packs labeled as follows: DIALYSIS ON/OFF BNDLE -NO DRSG, REF DYNDC3088

Recall: Z-1105-2024 · Reported March 6, 2024

Enforcement

Recall Number
Z-1105-2024
Event ID
93800
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 6, 2024
Initiation Date
November 15, 2023
Classification Date
February 24, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

MEDLINE Kits, trays, and packs labeled as follows: DIALYSIS ON/OFF BNDLE -NO DRSG, REF DYNDC3088

Reason

Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.

Code Info

UDI/DI 40193489884501 (case), 10193489884500 (unit), Lot Numbers: 21LBT672, 22BBG831

Distribution

Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama

Quantity

35 units