FDA Enforcement
Class II
Ongoing
cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number 09446125190
Recall: Z-1105-2023
·
Reported February 15, 2023
Enforcement
- Recall Number
- Z-1105-2023
- Event ID
- 91482
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Roche Molecular Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 15, 2023
- Initiation Date
- December 21, 2022
- Classification Date
- February 7, 2023
- Address
- 1080 Us Highway 202 S, Branchburg, NJ, 08876-3733, United States
Description
cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number 09446125190
Reason
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influenza as well as true diagnosis.
Code Info
UDI-DI 00875197006827 All in-date lots
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Quantity
9,948 OUS only