FDA Enforcement Class II Terminated

Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298

Recall: Z-1105-2020 · Reported February 12, 2020

Enforcement

Recall Number
Z-1105-2020
Event ID
83994
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Intuitive Surgical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2020
Initiation Date
December 4, 2018
Classification Date
February 6, 2020
Termination Date
August 24, 2022
Address
1266 Kifer Rd Bldg 101, N/A, Sunnyvale, CA, 94086-5304, United States

Description

Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298

Reason

This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.

Code Info

Serial Number/Manufacturing Lot: N11181003, N11181004

Distribution

US National Distribution Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming OUS:None

Quantity

286 Monopolar curved scissors