Product 7 consists of all products under product code HWC, and same usage: Item no: 47234801435 3.5MM CORT SCR X 14MM 47234801635 3.5MM CORT SCR X 16MM 47234806535 3.5MM CORT SCR X 65MM 47234807035 3.5MM CORT SCR X 70MM 47234807535 3.5MM CORT SCR X 75MM 47234808035 3.5MM CORT SCR X 80MM 47234808535 3.5MM CORT SCR X 85MM 47234809035 3.5MM CORT SCR X 90MM Product Usage: A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Enforcement
- Recall Number
- Z-1105-2016
- Event ID
- 72900
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Manufacturing B.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 23, 2016
- Initiation Date
- January 11, 2016
- Classification Date
- March 12, 2016
- Termination Date
- April 13, 2017
- Address
- Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, 00715, United States
Description
Product 7 consists of all products under product code HWC, and same usage: Item no: 47234801435 3.5MM CORT SCR X 14MM 47234801635 3.5MM CORT SCR X 16MM 47234806535 3.5MM CORT SCR X 65MM 47234807035 3.5MM CORT SCR X 70MM 47234807535 3.5MM CORT SCR X 75MM 47234808035 3.5MM CORT SCR X 80MM 47234808535 3.5MM CORT SCR X 85MM 47234809035 3.5MM CORT SCR X 90MM Product Usage: A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
lot no: 62230449 62081271 62133855 62159156 62047629 62127760 61903333 62261628 61830478 61830479 62138082 62230452 61800380 61800381
US Nationwide Distribution
109