FDA Enforcement Class II Ongoing

Integra Miltex Cryosolutions Cartridges 4 Pack (23.5g N2O each), Model: C-CA-23, Item: 33517;

Recall: Z-1104-2025 · Reported February 12, 2025

Enforcement

Recall Number
Z-1104-2025
Event ID
96067
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
New Medical Technologies Gmbh
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2025
Initiation Date
December 19, 2024
Classification Date
February 5, 2025
Address
Wankdorffeldstrasse 104, Bern, N/A, Switzerland

Description

Integra Miltex Cryosolutions Cartridges 4 Pack (23.5g N2O each), Model: C-CA-23, Item: 33517;

Reason

During the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.

Code Info

Model: C-CA-23; Item: 33517; UDI-DI: 10381780171966; Lots: C6741, D0122, D2560 and C6637.

Distribution

Worldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA and the country of Canada.

Quantity

3,352 units (838 packs of 4)