FDA Enforcement Class II Ongoing

Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals.

Recall: Z-1104-2021 · Reported February 17, 2021

Enforcement

Recall Number
Z-1104-2021
Event ID
87176
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Draegar Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 17, 2021
Initiation Date
December 23, 2020
Classification Date
February 10, 2021
Address
6 Tech Dr, N/A, Andover, MA, 01810-2434, United States

Description

Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals.

Reason

The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.

Code Info

Model Number - MS25510, MS25520, MS26372 Revisions 24 to 29

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lithuania, Malaysia, Mexico, Monaco, Namibia, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, United Kingdom, Utd.Arab.Emir., Vietnam.

Quantity

19802 systems (Domestic: 3412, Foreign 16,390)