FDA Enforcement Class II Terminated

Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804

Recall: Z-1104-2020 · Reported February 19, 2020

Enforcement

Recall Number
Z-1104-2020
Event ID
84436
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Dexcom Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
February 19, 2020
Initiation Date
November 19, 2019
Classification Date
February 7, 2020
Termination Date
August 22, 2023
Address
6340 Sequence Dr, San Diego, CA, 92121-4356, United States

Description

Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804

Reason

It was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days after installing the iOS application.

Code Info

Software Number: SW11677 Software Version: 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804

Distribution

US Nation-wide OUS: Austria; Belgium; Canada; Czechia; Denmark; Finland; Germany; Great Britain; Hong Kong; Ireland; Italy; Luxembourg; Netherlands; Norway; South Africa; Spain; Sweden, and Switzerland.

Quantity