FDA Enforcement Class II Terminated

Product 4 consists of all products under product code HWC, and same usage: Item no: 47482002603 2.0 X 26 CORT SCREW CRUCI 47482002803 2.0 X 28 CORT SCREW CRUCI 47482003203 2.0 X 32 CORT SCREW CRUCI 47482003403 2.0 X 34 CORT SCREW CRUCI 47482003603 2.0 X 36 CORT SCREW CRUCI 47482003803 2.0 X 38 CORT SCREW CRUCI 47482004003 2.0 X 40 CORT SCREW CRUCI 47482004203 2.0 X 42 CORT SCREW CRUCI 47482004403 2.0 X 44 CORT SCREW CRUCI 47482004603 2.0 X 46 CORT SCREW CRUCI 47482004803 2.0 X 48 CORT SCREW CRUCI 47482005003 2.0 X 50 CORT SCREW CRUCI 47482702000 2.7 X 20 CORT SCREW HEX Product Usage: A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Recall: Z-1102-2016 · Reported March 23, 2016

Enforcement

Recall Number
Z-1102-2016
Event ID
72900
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Manufacturing B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 23, 2016
Initiation Date
January 11, 2016
Classification Date
March 12, 2016
Termination Date
April 13, 2017
Address
Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, 00715, United States

Description

Product 4 consists of all products under product code HWC, and same usage: Item no: 47482002603 2.0 X 26 CORT SCREW CRUCI 47482002803 2.0 X 28 CORT SCREW CRUCI 47482003203 2.0 X 32 CORT SCREW CRUCI 47482003403 2.0 X 34 CORT SCREW CRUCI 47482003603 2.0 X 36 CORT SCREW CRUCI 47482003803 2.0 X 38 CORT SCREW CRUCI 47482004003 2.0 X 40 CORT SCREW CRUCI 47482004203 2.0 X 42 CORT SCREW CRUCI 47482004403 2.0 X 44 CORT SCREW CRUCI 47482004603 2.0 X 46 CORT SCREW CRUCI 47482004803 2.0 X 48 CORT SCREW CRUCI 47482005003 2.0 X 50 CORT SCREW CRUCI 47482702000 2.7 X 20 CORT SCREW HEX Product Usage: A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Reason

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Code Info

lot no.61951256 62230459 61907760 61907764 61926199 62261661 61926201 61907767 61907769 61907771 61909381 62230462 61907772 61196133 61907773 62020888 61920808

Distribution

US Nationwide Distribution

Quantity

154