FDA Enforcement Class II Ongoing

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Recall: Z-1101-2025 · Reported February 12, 2025

Enforcement

Recall Number
Z-1101-2025
Event ID
96047
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Nobel Biocare
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2025
Initiation Date
January 10, 2025
Classification Date
February 5, 2025
Address
Vastra Hamngatan 1, Goteborg, N/A, Sweden

Description

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Reason

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Code Info

Nobel Biocare N1 TiUltra TCC RP 4.0x 9mm Catalog Number: 300861 UDI-DI code: 07332747161731 Lot Numbers: 12253672 12253673 Nobel Biocare N1 TiUltra TCC RP 4.0x11mm Catalog Number: 300862 UDI-DI code: 07332747161748 Lot Numbers: 12253448 12254209 Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Catalog Number: 300863 UDI-DI code: 07332747161755 Lot Numbers: 12253620 12254213

Distribution

U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.

Quantity

1,558 units (90 units - U.S.)